Accuracy and clinical performance of a continuous intra-arterial blood-gas monitoring system during thoracoscopic surgery

被引:30
作者
Zollinger, A [1 ]
Spahn, DR [1 ]
Singer, T [1 ]
Zalunardo, MP [1 ]
Stoehr, S [1 ]
Weder, W [1 ]
Pasch, T [1 ]
机构
[1] UNIV ZURICH HOSP,DEPT SURG,CH-8091 ZURICH,SWITZERLAND
关键词
measurement techniques; gas exchange; equipment; blood-gas monitors; surgery; thoracic; ventilation; one-lung;
D O I
10.1093/bja/79.1.47
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Accuracy and performance of the only currently available intra-arterial blood-gas monitoring system (Paratrend 7, PT7) were assessed in 23 patients during thoracoscopic surgery using one-lung ventilation. Over a wide range of values for arterial Po-2 (6.1-61.1 kPa), Pco(2) (4.1-9.5 kPa) and pH (7.19-7.50), 138 arterial blood-gas values obtained by PT7 were compared with corresponding in vitro laboratory blood-gas measurements. We found good clinical performance with the PT7 and good agreement between PT7 values and in vitro measurements for arterial Po-2 (bias (1.96 SD)=0.38 (9.52) kPa), Pco(2) (0.31 (0.76) kPa) and pH (-0.017 (0.065)). Also, the bias for sequential changes between two consecutive times was not significantly different from the ideal value of 0. We conclude that the PT7 is helpful in monitoring patients during thoracoscopy.
引用
收藏
页码:47 / 52
页数:6
相关论文
共 34 条
[1]  
BARKER SJ, 1991, ANESTH ANALG, V73, P43
[2]   BLOOD-GAS ANALYSIS - EFFECT OF AIR BUBBLES IN SYRINGE AND DELAY IN ESTIMATION [J].
BISWAS, CK ;
RAMOS, JM ;
AGROYANNIS, B ;
KERR, DNS .
BRITISH MEDICAL JOURNAL, 1982, 284 (6320) :923-927
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Is continuous intra-arterial blood gas and pH monitoring justifiable? Editorial introduction [J].
Blitt, CD .
JOURNAL OF CLINICAL MONITORING, 1996, 12 (02) :177-177
[5]   TECHNOLOGY-ASSESSMENT OF ANESTHESIA MONITORS - PROBLEMS AND FUTURE-DIRECTIONS [J].
BYRICK, RJ ;
COHEN, MM .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1995, 42 (03) :234-239
[6]   THE EVALUATION OF A NEW INTRAVASCULAR BLOOD-GAS MONITORING-SYSTEM IN THE PIG [J].
CLUTTONBROCK, TH ;
FINK, S ;
MARKLE, D ;
LUTHRA, AJ ;
HENDRY, SP .
JOURNAL OF CLINICAL MONITORING, 1994, 10 (06) :387-391
[7]  
CORK RC, 1983, ANESTH ANALG, V62, P211
[8]   EFFECT OF SAMPLE DILUTIONS ON ARTERIAL BLOOD-GAS DETERMINATIONS [J].
DENNIS, RC ;
NG, R ;
YESTON, NS ;
STATLAND, B .
CRITICAL CARE MEDICINE, 1985, 13 (12) :1067-1068
[9]  
*ECRI, 1995, HLTH DEVICES, V24, P208
[10]  
EVERS W, 1972, ANESTH ANAL CURR RES, V51, P92