Inadequacy of motion correction algorithms in functional MRI: Role of susceptibility-induced artifacts

被引:63
作者
Wu, DH [1 ]
Lewin, JS [1 ]
Duerk, JL [1 ]
机构
[1] CASE WESTERN RESERVE UNIV, DEPT RADIOL, CLEVELAND, OH 44106 USA
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1997年 / 7卷 / 02期
关键词
functional MRI; artifact; susceptibility; image registration;
D O I
10.1002/jmri.1880070219
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Functional MR (fMR) has been useful in providing insight into the localization and detection of neural cortical activity. However, patient head motion is inevitable over the course of most fMR experiments. Although methods to align the control and activation fMR images may correct for some of this motional error, they will be incomplete in correcting for those that depend on spatial orientation. MR signal amplitude of structures that lie along air-tissue interfaces, for example, are sensitive to susceptibility errors introduced by their reorientation with respect to the main magnetic field. This analysis, using phantoms, cadaver brain, and volunteers, has shown that this reorientation between control and activation images can create regions of ''false activation'' that increase in area with rotation. Anatomic regions that are expected to be most affected by these susceptibility-induced magnetic field homogeneity variations are those adjacent to bone or air, such as the mesial and inferior temporal lobes and the inferior and anterior frontal lobes. Interpretation of ''activation'' in these areas on fMR studies must be performed with careful consideration of this important source of error.
引用
收藏
页码:365 / 370
页数:6
相关论文
共 12 条
[1]   PROCESSING STRATEGIES FOR TIME-COURSE DATA SETS IN FUNCTIONAL MRI OF THE HUMAN BRAIN [J].
BANDETTINI, PA ;
JESMANOWICZ, A ;
WONG, EC ;
HYDE, JS .
MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (02) :161-173
[2]   THE INTRAVASCULAR CONTRIBUTION TO FMRI SIGNAL CHANGE - MONTE-CARLO MODELING AND DIFFUSION-WEIGHTED STUDIES IN-VIVO [J].
BOXERMAN, JL ;
BANDETTINI, PA ;
KWONG, KK ;
BAKER, JR ;
DAVIS, TL ;
ROSEN, BR ;
WEISSKOFF, RM .
MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (01) :4-10
[3]   ARTIFACTS DUE TO STIMULUS CORRELATED MOTION IN FUNCTIONAL IMAGING OF THE BRAIN [J].
HAJNAL, JV ;
MYERS, R ;
OATRIDGE, A ;
SCHWIESO, JE ;
YOUNG, IR ;
BYDDER, GM .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (03) :283-291
[4]  
Haralick R. M., 1992, COMPUTER ROBOT VISIO
[5]   IDENTIFICATION OF VASCULAR STRUCTURES AS A MAJOR SOURCE OF SIGNAL CONTRAST IN HIGH-RESOLUTION 2D AND 3D FUNCTIONAL ACTIVATION IMAGING OF THE MOTOR CORTEX AT 1.5T - PRELIMINARY-RESULTS [J].
LAI, S ;
HOPKINS, AL ;
HAACKE, EM ;
LI, D ;
WASSERMAN, BA ;
BUCKLEY, P ;
FRIEDMAN, L ;
MELTZER, H ;
HEDERA, P ;
FRIEDLAND, R .
MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (03) :387-392
[6]  
LI S, 1995, 3 ANN M SOC MAGN RES, P731
[7]  
Ludeke K M, 1985, Magn Reson Imaging, V3, P329, DOI 10.1016/0730-725X(85)90397-2
[8]   FUNCTIONAL BRAIN MAPPING BY BLOOD OXYGENATION LEVEL-DEPENDENT CONTRAST MAGNETIC-RESONANCE-IMAGING - A COMPARISON OF SIGNAL CHARACTERISTICS WITH A BIOPHYSICAL MODEL [J].
OGAWA, S ;
MENON, RS ;
TANK, DW ;
KIM, SG ;
MERKLE, H ;
ELLERMANN, JM ;
UGURBIL, K .
BIOPHYSICAL JOURNAL, 1993, 64 (03) :803-812
[9]   OXYGENATION-SENSITIVE CONTRAST IN MAGNETIC-RESONANCE IMAGE OF RODENT BRAIN AT HIGH MAGNETIC-FIELDS [J].
OGAWA, S ;
LEE, TM ;
NAYAK, AS ;
GLYNN, P .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (01) :68-78
[10]   MICROSCOPIC SUSCEPTIBILITY VARIATION AND TRANSVERSE RELAXATION - THEORY AND EXPERIMENT [J].
WEISSKOFF, RM ;
ZUO, CS ;
BOXERMAN, JL ;
ROSEN, BR .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (06) :601-610