General and self-consistent method for the calibration of polarization modulators, polarimeters, and Mueller-matrix ellipsometers

被引:291
作者
Compain, E [1 ]
Poirier, S [1 ]
Drevillon, B [1 ]
机构
[1] Ecole Polytech, CNRS, UMR, Phys Interfaces & Couches Minces Lab, F-91128 Palaiseau, France
关键词
D O I
10.1364/AO.38.003490
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Calibration of polarization-state generators (PSG's), polarimeters, and Mueller-matrix ellipsometers (MME's) is an important factor in the practical use of these instruments. A new general procedure, the eigenvalue calibration method (ECM), is presented. It can calibrate any complete MME consisting of a PSG and a polarimeter that generate and measure, respectively, all the states of polarization of light. In the ECM, the PSG and the polarimeter are described by two 4 X 4 matrices W and A, and their 32 coefficients are determined from three or four measurements performed on reference samples. Those references are smooth isotropic samples and perfect linear polarizers. Their optical characteristics are unambiguously determined during the calibration from the eigenvalues of the measured matrices. The ECM does not require accurate alignment of the various optical elements and does not involve any first-order approximation. The ECM also displays an efficient error control capability that can be used to improve the MME behavior. The ECM is illustrated by an experimental calibration, at two wavelengths (458 and 633 nm), of a MME consisting of a coupled phase modulator associated with a prism division-of-amplitude polarimeter. (C) 1999 Optical Society of America. OCIS codes: 120.2130, 120.3930, 120.4640, 120.5410, 120.5820.
引用
收藏
页码:3490 / 3502
页数:13
相关论文
共 24 条
[1]  
[Anonymous], 1995, OPTICAL SCATTERING M
[3]   MUELLER-MATRIX MEASUREMENT USING THE 4-DETECTOR PHOTOPOLARIMETER [J].
AZZAM, RMA .
OPTICS LETTERS, 1986, 11 (05) :270-272
[4]   SIMULATION OF MECHANICAL ROTATION BY OPTICAL-ROTATION - APPLICATION TO DESIGN OF A NEW FOURIER PHOTOPOLARIMETER [J].
AZZAM, RMA .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1978, 68 (04) :518-521
[5]   ACCURATE CALIBRATION OF THE 4-DETECTOR PHOTOPOLARIMETER WITH IMPERFECT POLARIZING OPTICAL-ELEMENTS [J].
AZZAM, RMA ;
LOPEZ, AG .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1989, 6 (10) :1513-1521
[6]   ARRANGEMENT OF 4 PHOTODETECTORS FOR MEASURING THE STATE OF POLARIZATION OF LIGHT [J].
AZZAM, RMA .
OPTICS LETTERS, 1985, 10 (07) :309-311
[7]   DIVISION-OF-AMPLITUDE PHOTOPOLARIMETER (DOAP) FOR THE SIMULTANEOUS MEASUREMENT OF ALL 4 STOKES PARAMETERS OF LIGHT [J].
AZZAM, RMA .
OPTICA ACTA, 1982, 29 (05) :685-689
[8]   STOKES VECTORS, MUELLER MATRICES, AND POLARIZED SCATTERED-LIGHT [J].
BICKEL, WS ;
BAILEY, WM .
AMERICAN JOURNAL OF PHYSICS, 1985, 53 (05) :468-478
[9]   STATIC STOKES ELLIPSOMETER - GENERAL-ANALYSIS AND OPTIMIZATION [J].
BRUDZEWSKI, K .
JOURNAL OF MODERN OPTICS, 1991, 38 (05) :889-896
[10]   Deposition of SiO2 in integrated distributed electron cyclotron resonance microwave reactor [J].
Bulkin, P ;
Bertrand, N ;
Drevillon, B .
THIN SOLID FILMS, 1997, 296 (1-2) :66-68