Aneurysm clip testing for ferromagnetic properties: Clip variability issues

被引:42
作者
Kanal, E
Shellock, FG
Lewin, JS
机构
[1] UNIV SO CALIF,DEPT RADIOL,LOS ANGELES,CA 90089
[2] UNIV CLEVELAND HOSP,DEPT RADIOL,CLEVELAND,OH 44106
[3] CASE WESTERN RESERVE UNIV,CLEVELAND,OH 44106
关键词
aneurysm; cerebral; head; MR; magnetic resonance (MR); physics; safety; radiology and radiologists; iatrogenic injury;
D O I
10.1148/radiology.200.2.8685361
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To assess ferromagnetic properties of intracranial aneurysm clips reported to be nonferromagnetic, 1,765 Yasargil, 11 Sugita, and 15 Pemeczky aneurysm clips were studied for rotation or translation on plate glass in a 1.5-T MR imager. Sixty-three clips (52 Yasargil, 11 Perneczky) weakly reoriented along the static magnetic resonance (MR) field. These results confirm the need for standardized testing for ferromagnetic properties for implantable metallic devices.
引用
收藏
页码:576 / 578
页数:3
相关论文
共 16 条
[1]  
American Society for Testing and Materials, 1994, ANN BOOK ASTM STAND, V4.08
[2]  
BECKER RL, 1988, AM J NEURORADIOL, V9, P885
[3]   ANEURYSM CLIP MOTION DURING MAGNETIC-RESONANCE IMAGING - INVIVO EXPERIMENTAL-STUDY WITH METALLURGICAL FACTOR-ANALYSIS [J].
DUJOVNY, M ;
KOSSOVSKY, N ;
KOSSOWSKY, R ;
VALDIVIA, R ;
SUK, JS ;
DIAZ, FG ;
BERMAN, SK ;
CLEARY, W .
NEUROSURGERY, 1985, 17 (04) :543-548
[4]  
DUJOVNY M, 1995, 63 ANN M AM ASS NEUR, P132
[5]  
HOLTAS S, 1988, AM J NEURORADIOL, V9, P891
[6]   NEED FOR CAUTION DURING MR-IMAGING OF PATIENTS WITH ANEURYSM CLIPS [J].
JOHNSON, GC .
RADIOLOGY, 1993, 188 (01) :287-288
[7]   IMPORTANT CONSIDERATIONS IN MEASUREMENT OF ATTRACTIVE FORCE ON METALLIC IMPLANTS IN MR IMAGERS [J].
KAGETSU, NJ ;
LITT, AW .
RADIOLOGY, 1991, 179 (02) :505-508
[8]   SAFETY CONSIDERATIONS IN MR IMAGING [J].
KANAL, E ;
SHELLOCK, FG ;
TALAGALA, L .
RADIOLOGY, 1990, 176 (03) :593-606
[9]  
KANAL E, 1994, AM J NEURORADIOL, V15, P1394
[10]   MR-IMAGING OF PATIENTS WITH INTRACRANIAL ANEURYSM CLIPS [J].
KANAL, E ;
SHELLOCK, FG .
RADIOLOGY, 1993, 187 (03) :612-614