A validated GC-MS method for the determination and quantification of residual solvents in counterfeit tablets and capsules

被引:35
作者
Deconinck, E. [1 ]
Canfyn, M. [1 ]
Sacre, P. -Y. [1 ,2 ]
Baudewyns, S. [1 ]
Courselle, P. [1 ]
De Beer, J. O. [1 ]
机构
[1] Sci Inst Publ Hlth WIV ISP, Sect Med Prod, Div Food Med & Consumer Safety, B-1050 Brussels, Belgium
[2] Univ Liege, CIRM, Analyt Chem Lab, B-4000 Liege, Belgium
关键词
Residual solvents; Headspace; GC-MS; Counterfeit; Method validation; LINE CALIBRATION CURVES; DRUGS;
D O I
10.1016/j.jpba.2012.05.022
中图分类号
O65 [分析化学];
学科分类号
070302 [分析化学];
摘要
A fast headspace GC-MS method was developed and validated for the detection and quantification of residual solvents of all three ICH-classes in counterfeit tablets and capsules. The method was validated for W solvents, selected based on an initial screening of counterfeit medicinal products. The considered solvents were ethanol, 2-propanol, acetone, ethylacetate, chloroform, carbon tetrachloride, benzene, toluene, dichloromethane and ethylbenzene. The proposed method uses a Phenomenex 624 capillary column (60 m x 0.32 mm; 1.8 mu m film thickness) (Phenomenex, Torrance, USA) with an oven temperature program from 60 degrees C (held for 5 min) to 270 degrees C at 25 degrees C/min. 270 degrees C is held for 10 min. The total run time is 23.4 min. The obtained method was fully validated by applying the "total error" profile. Calibration lines for all components were linear within the studied ranges. The relative bias and the relative standard deviations for all components were smaller than 5%, the beta-expectation tolerance limits did not exceed the acceptance limits of +/- 10% and the relative expanded uncertainties were acceptable for all of the considered components. A method was obtained for the screening and quantification of residual solvents in counterfeit tablets and capsules, which will allow a fast screening of these products for the presence of residual solvents. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:64 / 70
页数:7
相关论文
共 20 条
[1]
[Anonymous], 2010, 35 US PHARM CONV INC
[2]
[Anonymous], 2005, 17025 EN ISOIEC
[3]
[Anonymous], 2011, JAP PHARM
[4]
Council of Europe, 2010, EUR PHARM 7 0
[5]
Improving quantitative gas chromatography-electron ionization mass spectrometry results using a modified ion source: Demonstration for a pharmaceutical application [J].
D'Autry, Ward ;
Wolfs, Kris ;
Hoogmartens, Jos ;
Adams, Erwin ;
Van Schepdael, Ann .
JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (26) :4034-4038
[6]
Mixed aqueous solutions as dilution media in the determination of residual solvents by static headspace gas chromatography [J].
D'Autry, Ward ;
Zheng, Chao ;
Wolfs, Kris ;
Yarramraju, Sitaramaraju ;
Hoogmartens, Jos ;
Van Schepdael, Ann ;
Adams, Erwin .
JOURNAL OF SEPARATION SCIENCE, 2011, 34 (11) :1299-1308
[7]
Liquid paraffin as new dilution medium for the analysis of high boiling point residual solvents with static headspace-gas chromatography [J].
D'Autry, Ward ;
Zheng, Chao ;
Bugalama, John ;
Wolfs, Kris ;
Hoogmartens, Jos ;
Adams, Erwin ;
Wang, Bochu ;
Van Schepdael, Ann .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2011, 55 (05) :1017-1023
[8]
Innovative development and validation of an HPLC/DAD method for the qualitative and quantitative determination of major cannabinoids in cannabis plant material [J].
De Backer, Benjamin ;
Debrus, Benjamin ;
Lebrun, Pierre ;
Theunis, Laetitia ;
Dubois, Nathalie ;
Decock, Lies ;
Verstraete, Alain ;
Hubert, Philippe ;
Charlier, Corinne .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2009, 877 (32) :4115-4124
[9]
Assessment of quality performance parameters for straight line calibration curves related to the spread of the abscissa values around their mean [J].
De Beer, Jacques O. ;
De Beer, Thomas R. ;
Goeyens, Leo .
ANALYTICA CHIMICA ACTA, 2007, 584 (01) :57-65
[10]
The quality coefficient as performance assessment parameter of straight line calibration curves in relationship with the number of calibration points [J].
De Beer, Jacques O. ;
Naert, Caroline ;
Deconinck, Eric .
ACCREDITATION AND QUALITY ASSURANCE, 2012, 17 (03) :265-274