A new penumbra generator for electron fields matching

被引:11
作者
Lachance, B
Tremblay, D
Pouliot, J
机构
[1] Ctr. Hosp. Univ. de Québec, Pavillon L'Hotel-Dieu de Quebec, Québec, Que.
关键词
electron beam; field matching; beam-edge modification; penumbra generator;
D O I
10.1118/1.597932
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Abutment of two or more electron fields to irradiate extended areas may lead to significant dose inhomogeneities in the junction region. This paper describes the geometric and dosimetric characteristics of a device developed to modify the penumbra of an electron beam and thereby improve the dose uniformity in the overlap region when fields are abutted. The device is a Lipowitz metal block placed on top of the electron applicator's insertion plate and positioned to stop part of the electron beam on the side of field abutment. The air-scattered electrons beyond the block increase the penumbra width from about 1.4 to 2.7-3.4 cm with an SSD of 100 cm. The modified penumbra is broad and almost linear at all depths for the 9 and 12 MeV electron beams used in this study. Film dosimetry was used to obtain beam profiles and isodose distributions of single modified beams and matched fields of 9 and 12 MeV as well as matched fields of both energies. Computer simulation was used to optimize the skin gap to be used and to quantify the dose uniformity as a function of the field separation for both modified and nonmodified beams. Results are presented for various field configurations. Without the penumbra generator, lateral setup errors of 2-3 mm may introduce dose variations of 20% or more in the junction region. Similar setup errors cause less than 5% dose variations when the penumbra generator is used to match the fields. The potential of the technique for the irradiation of curved surfaces is presented. A possible method for implementing the modified penumbra into a conventional treatment planning system is evaluated. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:485 / 495
页数:11
相关论文
共 16 条
[1]  
ALMOND PR, 1976, CLIN APPL ELECT BEAM
[2]   ADJACENT FIELDS OF HIGH-ENERGY X-RAYS AND ELECTRONS - FLAT SURFACES [J].
BAGNE, F .
PHYSICS IN MEDICINE AND BIOLOGY, 1978, 23 (06) :1186-1191
[3]   AN ELECTRON-BEAM PSEUDO-ARC TECHNIQUE FOR IRRADIATION OF LARGE AREAS OF CHEST WALL AND OTHER CURVED SURFACES [J].
BOYER, AL ;
FULLERTON, GD ;
MIRA, JG .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1982, 8 (11) :1969-1974
[4]  
Feygelman V, 1994, Med Dosim, V19, P23
[5]   ABUTMENT OF HIGH-ENERGY ELECTRON FIELDS [J].
HARMS, WB ;
PURDY, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 20 (04) :853-858
[6]   A BEAM-EDGE MODIFIER FOR ABUTTING ELECTRON FIELDS [J].
KALEND, AM ;
ZWICKER, RD ;
WU, A ;
STERNICK, ES .
MEDICAL PHYSICS, 1985, 12 (06) :793-798
[7]   FIELD MATCHING OF ELECTRON-BEAMS USING PLASTIC WEDGE PENUMBRA GENERATORS [J].
KURUP, RG ;
WANG, S ;
GLASGOW, GP .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01) :145-153
[8]   DEMONSTRATION OF RELATIVELY NEW ELECTRON DOSIMETRY MEASUREMENT TECHNIQUES ON THE MEVATRON-80 [J].
MEYER, JA ;
PALTA, JR ;
HOGSTROM, KR .
MEDICAL PHYSICS, 1984, 11 (05) :670-677
[9]  
Norris M, 1991, Med Dosim, V16, P153
[10]   THE REVERSE HOCKEY STICK TECHNIQUE - POSTMASTECTOMY RADIATION-THERAPY FOR BREAST-CANCER PATIENTS WITH LOCALLY ADVANCED TUMOR PRESENTATION OR EXTENSIVE LOCO-REGIONAL RECURRENCE [J].
PEZNER, RD ;
LIPSETT, JA ;
FORELL, B ;
VORA, NL ;
DESAI, KR ;
WONG, JY ;
LUK, KH .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 17 (01) :191-197