Evolution of atrial signals from a single lead VDD pacemaker

被引:13
作者
Papouchado, M [1 ]
Crick, JCP [1 ]
机构
[1] BRISTOL HEART INST,BRISTOL,AVON,ENGLAND
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1996年 / 19卷 / 11期
关键词
atrioventricular synchronous pacing; atrial signals;
D O I
10.1111/j.1540-8159.1996.tb03222.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The atrial sensing capabilities of a new single pass lead VDD pacing system (Pacesetter AddVent) were assessed in a prospective multicenter study of 101 implants during the period July 1994 through March 1996. The pacing lead (Pacesetter AV Plus) has (I unique quadripolar Gin-line connector and uses a pair of ring electrodes with an interelectrode spacing of 12 mm for atrial sensing. The mean age of the patients (51 men) was 73 years (range 19-91). Seventy-five patients had complete heart block; the others had 2:1 AV block. Wide variations were found in signal amplitude: mean P wave amplitude, measured over four cycles in the supine position, was 2.4 +/- 1.9 mV at implant, dropping to 1.9 +/- 1.7mV predischarge, and remaining constant at follow-up but with a narrower range. Holter monitoring was undertaken in 24 patients, with a total of 550 monitored hours; Mean AV synchrony was 98.2% +/- 4.6% (excluding premature ventricular contractions), with 20 patients (83%) showing > 99% AV synchrony, with atrial sensing at 0.1 mV where needed. No oversensing was observed in any patient. There was a low incidence of atrial fibrillation (2%) and sinus bradycardia (0%). The findings show that the range of atrial signals, although wide initially, converges over the first year and remains adequate for reliable AV synchronous pacing.
引用
收藏
页码:1772 / 1776
页数:5
相关论文
共 9 条
[1]   LONG-TERM RELIABILITY OF SINGLE-LEAD ATRIAL SYNCHRONOUS PACING SYSTEMS USING CLOSELY SPACED ATRIAL DIPOLES - 5-YEAR EXPERIENCE [J].
ANSANI, L ;
PERCOCO, GF ;
GUARDIGLI, G ;
TOSELLI, T ;
ANTONIOLI, GE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11) :1865-1869
[2]   SINGLE-LEAD ATRIAL SYNCHRONOUS VENTRICULAR PACING - A DREAM COME TRUE [J].
ANTONIOLI, GE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (09) :1531-1547
[3]  
CLARKE M, 1991, BRIT HEART J, V66, P185
[4]   EUROPEAN MULTICENTER PROSPECTIVE FOLLOW-UP-STUDY OF 1,002 IMPLANTS OF A SINGLE LEAD VDD PACING SYSTEM [J].
CRICK, JCP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11) :1742-1744
[5]  
CRICK JCP, 1995, PACE, V18, P1785
[6]   LONG-TERM STABILITY OF P-WAVE SENSING IN SINGLE-LEAD VDDR PACING - CLINICAL VERSUS SUBCLINICAL ATRIAL UNDERSENSING [J].
LAU, CP ;
TAI, YT ;
LEUNG, SK ;
LEUNG, WH ;
CHUNG, FLW ;
LEE, ISF .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11) :1849-1853
[7]   CLINICAL-EVALUATION OF A NEW SINGLE-PASS LEAD VDD PACING SYSTEM [J].
OVSYSHCHER, IE ;
KATZ, A ;
BONDY, C .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11) :1859-1864
[8]  
PAPOUCHADO M, 1995, EUROPACE 95, P571
[9]  
TABORSKI M, 1993, PACE, V16, P1937