Immunoassay for thyroxine (T4) in serum using capillary electrophoresis and micromachined devices

被引:90
作者
Schmalzing, D [1 ]
Koutny, LB [1 ]
Taylor, TA [1 ]
Nashabeh, W [1 ]
Fuchs, M [1 ]
机构
[1] PERSEPT BIOSYST INC, FRAMINGHAM, MA 01701 USA
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 1997年 / 697卷 / 1-2期
关键词
thyroxine;
D O I
10.1016/S0378-4347(97)00256-9
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
As part of our ongoing effort to develop electrophoretic assay technology for clinical diagnostics, we describe a competitive immunoassay for the determination of serum thyroxine (T4) based on electrophoresis and laser induced fluorescence (LIF). Measurements of total T4 are useful for the clinical evaluation of thyroid function. A fluorescein thyroxine conjugate was utilized in conjunction with a polyclonal antibody preparation as assay reagents. Capillary electrophoresis (CE) conditions tolerant of the direct injection of serum without extraction or other sample preparation steps were developed and used for quantitation of total T4 in serum. We have been exploring the use of micromachined devices with arrays of channels for high assay throughput. Our assay protocol was carried in a microchip format. The results illustrate that gains in speed can be additionally achieved, with the electrophoretic separation of free from bound labelled T4 being performed in about 15 s for serum samples. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:175 / 180
页数:6
相关论文
共 16 条
[1]  
Blais Burton W., 1994, Food and Agricultural Immunology, V6, P409, DOI 10.1080/09540109409354853
[2]  
CHEN FTA, 1994, CLIN CHEM, V40, P1819
[3]  
HARRISON DJ, 1996, SOL STAT SENS ACT WO
[4]   EFFECTS OF INJECTION SCHEMES AND COLUMN GEOMETRY ON THE PERFORMANCE OF MICROCHIP ELECTROPHORESIS DEVICES [J].
JACOBSON, SC ;
HERGENRODER, R ;
KOUTNY, LB ;
WARMACK, RJ ;
RAMSEY, JM .
ANALYTICAL CHEMISTRY, 1994, 66 (07) :1107-1113
[5]   Microchip electrophoretic immunoassay for serum cortisol [J].
Koutny, LB ;
Schmalzing, D ;
Taylor, TA ;
Fuchs, M .
ANALYTICAL CHEMISTRY, 1996, 68 (01) :18-22
[6]  
KOUTNY LB, 1996, HPCE 96 ORL FLOR US
[7]   PLANAR CHIPS TECHNOLOGY FOR MINIATURIZATION AND INTEGRATION OF SEPARATION TECHNIQUES INTO MONITORING SYSTEMS - CAPILLARY ELECTROPHORESIS ON A CHIP [J].
MANZ, A ;
HARRISON, DJ ;
VERPOORTE, EMJ ;
FETTINGER, JC ;
PAULUS, A ;
LUDI, H ;
WIDMER, HM .
JOURNAL OF CHROMATOGRAPHY, 1992, 593 (1-2) :253-258
[8]   CAPILLARY ARRAY ELECTROPHORESIS - AN APPROACH TO HIGH-SPEED, HIGH-THROUGHPUT DNA SEQUENCING [J].
MATHIES, RA ;
HUANG, XC .
NATURE, 1992, 359 (6391) :167-169
[9]   FLUORESCEIN ISOTHIOCYANATE-LABELED PROTEIN-G AS AN AFFINITY LIGAND IN AFFINITY IMMUNOCAPILLARY ELECTROPHORESIS WITH FLUORESCENCE DETECTION [J].
REIF, OW ;
LAUSCH, R ;
SCHEPER, T ;
FREITAG, R .
ANALYTICAL CHEMISTRY, 1994, 66 (22) :4027-4033
[10]   CHARACTERIZATION AND PERFORMANCE OF A NEUTRAL HYDROPHILIC COATING FOR THE CAPILLARY ELECTROPHORETIC SEPARATION OF BIOPOLYMERS [J].
SCHMALZING, D ;
PIGGEE, CA ;
FORET, F ;
CARRILHO, E ;
KARGER, BL .
JOURNAL OF CHROMATOGRAPHY A, 1993, 652 (01) :149-159