Cup inclination angle of greater than 50 degrees increases whole blood concentrations of cobalt and chromium ions after metal-on-metal hip resurfacing

被引:113
作者
Hart, A. J. [1 ]
Buddhdev, P. [2 ]
Winship, P. [3 ]
Faria, N. [3 ]
Powell, J. J. [3 ]
Skinner, J. A. [2 ]
机构
[1] Charing Cross Hosp, Imperial Coll Healthcare NHS Trust, Dept Musculoskeletal Surg, London, England
[2] UCL, Royal Natl Orthopaed Hosp, London, England
[3] MRC Human Nutr Res Ctr, Cambridge, England
基金
英国医学研究理事会;
关键词
Metal on metal; Hip arthroplasty; Hip replacement; Hip resurfacing; Metal ions; In vivo wear rate; Inclination angle; Component position; Biomarker;
D O I
10.5301/HIP.2008.948
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A cup inclination angle greater than 45 degrees is associated with increased wear rates of metal on polyethylene (MOP) hip replacements. The same maybe true for metal on metal (MOM) hips yet this has not been clearly shown. We measured the acetabular inclination angle from plain radiographs, and whole blood metal ion levels using Inductively Coupled Plasma Mass Spectrometry of 26 patients (mean Harris Hip Score 94 and mean time post op of 22 months) with Birmingham Hip Resurfacings. We identified a threshold level of 50 degrees cup inclination. Below this threshold, the mean whole blood cobalt and chromium were 1.6ppb and 1.88ppb respectively, above this threshold, the mean blood cobalt and chromium were 4.45ppb and 4.3ppb respectively. These differences were significant cobalt (p < 0.01) and chromium (p=0.01). All patients above the threshold had metal levels greater than any of the patients below the threshold. For 14 patients, who returned one year later for a repeat blood metal level measurement, cobalt and chromium levels were very similar The effect of an acetabular inclination angle of greater than 50 degrees on wear rates of MOM hips, as measured through blood metal ion levels, appears to be similar to that seen with MOP hips. Additionally, our new analytical methods may allow blood metal levels to be used as a realistic biomarker of in vivo wear rate of MOM hips. The implication is that metal levels can be minimised with optimal orientation of the acetabular component. (Hip International 2008; 18: 212-9)
引用
收藏
页码:212 / 219
页数:8
相关论文
共 41 条
[1]
METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS [J].
AGINS, HJ ;
ALCOCK, NW ;
BANSAL, M ;
SALVATI, EA ;
WILSON, PD ;
PELLICCI, PM ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :347-356
[2]
ANGADJI A, 2008, T 54 ANN M ORTH RES
[3]
How do serum cobalt and chromium levels change after metal-on-metal hip resurfacing? [J].
Back, DL ;
Young, AA ;
Shimmin, AJ .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (438) :177-181
[4]
Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty [J].
Berry, DJ ;
Von Knoch, M ;
Schleck, CD ;
Harmsen, WS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (11) :2456-2463
[5]
Metal-on-metal bearing with THA and chronic renal failure. [J].
Brodner, V ;
Grohs, JG ;
Bitzan, P ;
Meisinger, V ;
Kovarik, J ;
Kotz, R .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2000, 138 (05) :425-429
[6]
Serum cobalt levels after metal-on-metal total hip arthroplasty [J].
Brodner, W ;
Bitzan, P ;
Meisinger, V ;
Kaider, A ;
Gottsauner-Wolf, F ;
Kotz, R .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (11) :2168-2173
[7]
Cup inclination and serum concentration of cobalt and chromium after metal-on-metal total hip arthroplasty [J].
Brodner, W ;
Grübl, A ;
Jankovsky, R ;
Meisinger, V ;
Lehr, S ;
Gottsauner-Wolf, F .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :66-70
[8]
Chromosomal changes after surgery for joint replacement [J].
Case, CP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (08) :1093-1095
[9]
Case CP, 2001, CLIN CHEM, V47, P275
[10]
Comparison of in vitro with in vivo characteristics of wear particles from metal-metal hip implants [J].
Catelas, I ;
Medley, JB ;
Campbell, PA ;
Huk, OL ;
Bobyn, JD .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2004, 70B (02) :167-178