Hemodynamics and early clinical performance of the St. Jude Medical Regent mechanical aortic valve

被引:43
作者
Bach, DS [1 ]
Sakwa, MP
Goldbach, M
Petracek, MR
Emery, RW
Mohr, FW
机构
[1] Univ Michigan, Dept Med, Div Cardiol, Ann Arbor, MI 48109 USA
[2] William Beaumont Hosp, Dept Surg, Sect Cardiothorac Surg, Royal Oak, MI 48072 USA
[3] London Hlth Sci Ctr, Dept Surg, Sect Cardiothorac Surg, London, ON, Canada
[4] St Thomas Hosp, Dept Surg, Sect Cardiothorac Surg, Nashville, TN USA
[5] Univ Leipzig, Herzzentrum, Dept Surg, Sect Cardiothorac Surg, Leipzig, Germany
关键词
D O I
10.1016/S0003-4975(02)04034-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. The St. Jude Medical Regent valve is the next-generation bileaflet aortic prosthesis, modified from the currently marketed St. Jude Medical mechanical valve to achieve a larger geometric orifice without changing the existing design of the pivot mechanism or blood-contact surface areas. The present study reports the hemodynamic and early clinical results of an ongoing multicenter trial investigating the performance of the Regent valve. Methods. Between July 1998 and July 2001, 361 patients at 17 centers in North America and Europe underwent implantation of a Regent mechanical aortic valve prosthesis. Clinical status was prospectively recorded, and echocardiography with Doppler was performed at discharge and at 2 months, 6 months, 1 year, and 2 years after operation. Results. Follow-up to date is 300 patient-years (average, 0.8 +/- 0.7 years per patient; range, 0.0 to 2.7 years). There were low rates of clinical adverse events. Mean gradient at 6 months was 9.7 +/- 5.3 mm Hg, 7.6 +/- 5.2 mm Hg,6.3 +/- 3.7 mm Hg, 5.8 +/- 3.4 mm Hg, and 4.0 +/- 2.6 mm Hg, respectively, for 19-mm, 21-mm, 23-mm, 25-mm, and 27-mm valves; effective orifice area was 1.6 +/- 0.4 cm(2), 2.0 +/- +/- 0.7 cm(2), 2.2 +/- 0.9 cm(2), 2.5 +/- 0.9 cm(2), and 3.6 +/- 1.3 cm(2), respectively. Indexed effective orifice area was equal to or greater than 1.0 cm(2)/m(2) for all valve sizes. Left ventricular mass index decreased significantly between early postoperative (165.9 +/- 57.1 g/m(2)) and 6-month follow-up (137.9 +/- 41.0 g/m(2); Delta = -28.0 +/- 49.1 g/m(2); p < 0.0001). Conclusions. The St. Jude Medical Regent aortic valve has excellent hemodynamics and early clinical results, with rapid and significant left ventricular mass regression. Long-term clinical assessment is ongoing.
引用
收藏
页码:2003 / 2009
页数:7
相关论文
共 9 条
[1]   Hemodynamics and left ventricular mass regression following implantation of the Toronto SPV stentless porcine valve [J].
Bach, DS ;
David, T ;
Yacoub, M ;
Pepper, J ;
Goldman, B ;
Wood, J ;
Verrier, F ;
Petracek, M ;
Aldrete, V ;
Rosenbloom, M ;
Azar, H ;
Rakowski, H .
AMERICAN JOURNAL OF CARDIOLOGY, 1998, 82 (10) :1214-1219
[2]   LONG-TERM RESULTS OF VALVE-REPLACEMENT WITH THE ST-JUDE MEDICAL PROSTHESIS [J].
BAUDET, EM ;
PUEL, V ;
MCBRIDE, JT ;
GRIMAUD, JP ;
ROQUES, F ;
CLERC, F ;
ROQUES, X ;
LABORDE, N .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1995, 109 (05) :858-870
[3]   ECHOCARDIOGRAPHIC ASSESSMENT OF LEFT-VENTRICULAR HYPERTROPHY - COMPARISON TO NECROPSY FINDINGS [J].
DEVEREUX, RB ;
ALONSO, DR ;
LUTAS, EM ;
GOTTLIEB, GJ ;
CAMPO, E ;
SACHS, I ;
REICHEK, N .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (06) :450-458
[4]   Guidelines for reporting morbidity and mortality after cardiac valvular operations [J].
Edmunds, LH ;
Clark, RE ;
Cohn, LH ;
Grunkemeier, GL ;
Miller, DC ;
Weisel, RD .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1996, 112 (03) :708-711
[5]  
Feigenbaum H., 1994, ECHOCARDIOGR-J CARD, P181
[6]  
Otto CM, 2000, TXB CLIN ECHOCARDIOG, P310
[7]   EVALUATION OF AORTIC-INSUFFICIENCY BY DOPPLER COLOR FLOW MAPPING [J].
PERRY, GJ ;
HELMCKE, F ;
NANDA, NC ;
BYARD, C ;
SOTO, B .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 9 (04) :952-959
[8]   Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention [J].
Pibarot, P ;
Dumesnil, JG .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 36 (04) :1131-1141
[9]   PROBLEM OF VALVE PROSTHESIS PATIENT MISMATCH [J].
RAHIMTOOLA, SH .
CIRCULATION, 1978, 58 (01) :20-24