Validation of the RT3 triaxial accelerometer for the assessment of physical activity

被引:259
作者
Rowlands, AV [1 ]
Thomas, PWM [1 ]
Eston, RG [1 ]
Topping, R [1 ]
机构
[1] Univ Wales, Sch Sport Hlth & Exercise Sci, Bangor LL57 2PX, Gwynedd, Wales
关键词
motion sensor; oxygen consumption; habitual physical activity; energy expenditure;
D O I
10.1249/01.MSS.0000117158.14542.E7
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
Purpose: The aims of this study were to assess and compare the validity of the RT3 accelerometer for the assessment of physical activity in boys and men, to compare RT3 and Tritrac accelerometer counts, and to determine count cut-off values for moderate (greater than or equal to3 < 6 METs) and vigorous (greater than or equal to6 METs) activity. Methods: Nineteen boys (age: 9.5 +/- 0.8 yr) and 15 men (age: 20.7 +/- 1.4 yr) walked and ran on a treadmill, kicked a ball to and fro, played hopscotch, and sat quietly. An RT3 was worn on the right hip; boys also wore a Tritrac on the left hip. Oxygen consumption was expressed as a ratio of body mass raised to the power of 0.75 (S(V) over dot O-2). Results: RT3 counts correlated significantly with S(V) over dot O-2 in boys (r = 0.87, P < 0.01) and men (r = 0.85, P < 0.01). However, during treadmill activities, RT3 counts were significantly higher for boys (P < 0.05). RT3 counts corresponding to "moderate" and "vigorous" activity were similar for boys and men for all activities (moderate = 970.2 for boys and 984.0 for men-, vigorous = 2333.0 for boys and 2340.8 for men) but approximately 400 counts lower in men when only treadmill activities were considered. Tritrac counts correlated significantly with S(V) over dot O-2 in boys (r = 0.87, P < 0.01), but were significantly lower than RT3 counts across most activities (P < 0.05). Conclusions: The RT3 accelerometer is a good measure of physical activity for boys and men. However, moderate and vigorous intensity count thresholds differ for boys and men when the predominant activities are walking and running. RT3 counts are significantly higher than Tritrac counts for a number of activities. These findings have implications when comparing activity counts between studies using the different instruments.
引用
收藏
页码:518 / 524
页数:7
相关论文
共 22 条
[1]   Validity and reliability issues in objective monitoring of physical activity [J].
Bassett, DR .
RESEARCH QUARTERLY FOR EXERCISE AND SPORT, 2000, 71 (02) :S30-S36
[2]   Validity of four motion sensors in measuring moderate intensity physical activity [J].
Bassett, DR ;
Ainsworth, BE ;
Swartz, AM ;
Strath, SJ ;
O'Brien, WL ;
King, GA .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2000, 32 (09) :S471-S480
[3]  
BOUTEN CV, 1994, MED SCI SPORT EXER, V26, P1516
[4]  
Dale D, 2002, PHYS ACTIVITY ASSESS, P19
[5]   Validity of heart rate, pedometry, and accelerometry for predicting the energy cost of children's activities [J].
Eston, RG ;
Rowlands, AV ;
Ingledew, DK .
JOURNAL OF APPLIED PHYSIOLOGY, 1998, 84 (01) :362-371
[6]  
ESTON RG, 1993, KINANTHROPOMETRY, V4, P236
[7]   Validity of heart rate, pedometry, and accelerometry for estimating the energy cost of activity in Hong Kong Chinese boys [J].
Louie, L ;
Eston, RG ;
Rowlands, AV ;
Tong, KK ;
Ingledew, DK ;
Fu, FH .
PEDIATRIC EXERCISE SCIENCE, 1999, 11 (03) :229-239
[8]  
MEIJER GA, 1989, MED SCI SPORT EXER, V21, P343
[9]  
MELANSON EL, 1995, MED SCI SPORTS EXERC, V27, P1071
[10]   COMPARING CORRELATED CORRELATION-COEFFICIENTS [J].
MENG, XL ;
ROSENTHAL, R ;
RUBIN, DB .