Fixation strength of meniscal repair devices

被引:27
作者
Bellemans, J [1 ]
Vandenneucker, H [1 ]
Labey, L [1 ]
Van Audekercke, R [1 ]
机构
[1] Katholieke Univ Leuven, Univ Hosp Pellenberg, Dept Orthopaed Surg, B-3212 Lubbeck, Pellenberg, Belgium
关键词
meniscus; repair; devices;
D O I
10.1016/S0968-0160(01)00126-0
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The aim of this study was to measure and compare the ultimate failure strengths and cyclic fatigue strengths of currently available meniscal suturing devices. No statistically significant difference in failure load was found between a vertical loop suture (mean 46.3 N) a horizontal mattress suture (52.5 N), the T-Ex Device(R) (47.5 N), and the 16-mm (39.2 N) and 13-mm (32.8 N) Bionix Arrow(R). Statistically inferior results were seen with the 10-mm Bionix Arrow(R) (18.8 N), the S.D. Sorb Stapler(R) (4.3 N) and the 12-mm Arthrex Meniscal Dart(R) (10.5 N) (P < 0.01). The Mitek Meniscal Repair System(R) (28.1 N) performed intermediate, with significantly better results than the S.D. Sorb Stapler(R) and the 12-mm Arthrex Dart(R), but significantly worse than the vertical and horizontal sutures, the T-fix(R) and Bionix 16-mm Device(R) (P < 0.01). Cyclic fatigue strength was significantly less for the 10-mm Bionix Arrow(R), the S.D. Sorb Stapler(R), and the 12-mm Arthrex Meniscal Dart(R) (P < 0.01) compared to all other devices. We conclude that the 13- and 16-mm Bionix Arrow(R) and the T-fix Device(R) have comparable ultimate failure strengths and cyclic fatigue strengths to conventional meniscal suturing techniques. The 10-mm Bionix Arrow(R), S.D. Sorb Stapler(R), and the 12-mm Arthrex Meniscal Dart(R), however, have far inferior failure and cyclic fatigue strengths and their clinical application should be questioned. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:11 / 14
页数:4
相关论文
共 10 条
[1]   The arrow versus horizontal suture in arthroscopic meniscus repair - A prospective randomized study with arthroscopic evaluation [J].
Albrecht-Olsen, P ;
Kristensen, G ;
Burgaard, P ;
Joergensen, U ;
Toerholm, C .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 1999, 7 (05) :268-273
[2]   Failure strength of a new meniscus arrow repair technique: Biomechanical comparison with horizontal suture [J].
AlbrechtOlsen, P ;
Lind, T ;
Kristensen, G ;
Falkenberg, B .
ARTHROSCOPY, 1997, 13 (02) :183-187
[3]   Meniscal repair devices [J].
Barber, FA ;
Herbert, MA .
ARTHROSCOPY, 2000, 16 (06) :613-618
[4]   Preliminary results of the T-fix endoscopic meniscus repair technique in an anterior cruciate ligament reconstruction population [J].
Barrett, GR ;
Treacy, SH ;
Ruff, CG .
ARTHROSCOPY, 1997, 13 (02) :218-223
[5]   Pull-out strength and stiffness of meniscal repair using absorbable arrows or Ti-Cron vertical and horizontal loop sutures [J].
Boenisch, UW ;
Faber, KJ ;
Ciarelli, M ;
Steadman, JR ;
Arnoczky, SP .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1999, 27 (05) :626-631
[6]   Failure strengths of suture versus biodegradable arrow for meniscal repair: An in vitro study [J].
Dervin, GF ;
Downing, KJW ;
Keene, GCR ;
McBride, DG .
ARTHROSCOPY, 1997, 13 (03) :296-300
[7]   T-Fix anchor sutures for arthroscopic meniscal repair [J].
Escalas F. ;
Quadras J. ;
Cáceres E. ;
Benaddi J. .
Knee Surgery, Sports Traumatology, Arthroscopy, 1997, 5 (2) :72-76
[8]   Biofix resorbable meniscus arrow for meniscal ruptures:: results of a 1-year follow-up [J].
Hürel, C ;
Mertens, F ;
Verdonk, R .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2000, 8 (01) :46-52
[9]   Meniscal fixation with an absorbable staple: An experimental study in dogs [J].
Koukoubis T.D. ;
Glisson R.R. ;
Feagin Jr. J.A. ;
Seaber A.V. ;
Schenkman D. ;
Korompilias A.V. ;
Stahl D.L. .
Knee Surgery, Sports Traumatology, Arthroscopy, 1997, 5 (1) :22-30
[10]   Biomechanical test comparing the load to failure of the biodegradable meniscus arrow versus meniscal suture [J].
Song, EK ;
Lee, KB .
ARTHROSCOPY, 1999, 15 (07) :726-732