A double-blind study of capacitively coupled electrical stimulation as an adjunct to lumbar spinal fusions

被引:79
作者
Goodwin, CB
Brighton, CT
Guyer, RD
Johnson, JR
Light, KI
Yuan, HA
机构
[1] Hosp Special Surg, New York, NY 10021 USA
[2] Univ Penn, Dept Orthopaed Surg, Philadelphia, PA 19104 USA
[3] Texas Back Inst Res Fdn, Plano, TX USA
[4] Univ Louisville, Sch Med, Dept Orthopaed Surg, Louisville, KY 40292 USA
[5] St Francis Mem Hosp, San Francisco Spine Ctr, San Francisco, CA 94109 USA
[6] SUNY Syracuse, Hlth Sci Ctr, Syracuse, NY 13210 USA
关键词
electrical stimulation; lumbar fusion; pedicle screws; posterolateral lumbar fusion; randomized trial;
D O I
10.1097/00007632-199907010-00013
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. A randomized double-blind prospective comparison with a placebo control. This report of the results is the first in an ongoing study. Objectives. To evaluate the effect of noninvasive capacitively coupled electrical stimulation on the success rate of lumbar spine fusion surgery, and to compare active with placebo stimulators as adjuncts to contemporary fusion techniques. Summary of Background Data. Previous studies have established the effectiveness of direct current and electromagnetic field stimulation as adjuncts for some forms of spinal fusion. None of the previous placebo-controlled studies on external bone stimulation included posterolateral fusion techniques, and most were conducted with prior generations of internal fixation hardware. Methods. The investigation was conducted by 28 U.S. surgeons. Patients with a primary diagnosis of degenerative disc disease with or without other degenerative changes were selected. The study protocol defined success as a clinical outcome rated as excellent or good and a fusion documented as solid by both the investigator and dhe blinded independent radiologist. Disagreements on radiographic success were resolved by a second blinded independent reviewer. Results. For the 179 patients who completed treatment and evaluation, the overall protocol success rate (both clinical and radiographic results rated as successes) was 84.7% for the active patients and 64.9% for the placebo I-patients. This difference is highly significant according to the Yates corrected chi-square test (P = 0.0043). Best improvements in patient outcomes (20% or greater success rate) occurred when active stimulation was used in conjunction with posterolateral fusion (P = 0.006) and when internal fixation also was incorporated (P = 0.013). Discussion. This study was consistent in that active stimulation improved results for each stratification, although some strata had insufficient numbers of patients for the-results to have statistical significance, Improved success rates when capacitively coupled stimulation is added to internal fixation are hypothesized to result from overcoming the biochemical effects of stress shielding. Conclusions. Capacitively coupled stimulation is an effective adjunct to primary spine fusion, especially for patients with posterolateral fusion and those with internal fixation.
引用
收藏
页码:1349 / 1356
页数:8
相关论文
共 44 条
[1]  
ARCHER C, 1981, T BIOELECTRIC REPAIR, V1, P1
[2]   STIMULATION OF SKELETAL-DERIVED CELL-CULTURES BY DIFFERENT ELECTRIC-FIELD INTENSITIES IS CELL-SPECIFIC [J].
BINDERMAN, I ;
SOMJEN, D ;
SHIMSHONI, Z ;
LEVY, J ;
FISCHLER, H ;
KORENSTEIN, R .
BIOCHIMICA ET BIOPHYSICA ACTA, 1985, 844 (03) :273-279
[3]   TREATMENT OF DENERVATION DISUSE OSTEOPOROSIS IN THE RAT WITH A CAPACITIVELY COUPLED ELECTRICAL SIGNAL - EFFECTS ON BONE-FORMATION AND BONE-RESORPTION [J].
BRIGHTON, CT ;
TADDUNI, GT ;
GOLL, SR ;
POLLACK, SR .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1988, 6 (05) :676-684
[4]   TREATMENT OF RECALCITRANT NON-UNION WITH A CAPACITIVELY COUPLED ELECTRICAL-FIELD - A PRELIMINARY-REPORT [J].
BRIGHTON, CT ;
POLLACK, SR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1985, 67A (04) :577-585
[5]   FRACTURE-HEALING IN THE RABBIT FIBULA WHEN SUBJECTED TO VARIOUS CAPACITIVELY COUPLED ELECTRICAL FIELDS [J].
BRIGHTON, CT ;
HOZACK, WJ ;
BRAGER, MD ;
WINDSOR, RE ;
POLLACK, SR ;
VRESLOVIC, EJ ;
KOTWICK, JE .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1985, 3 (03) :331-340
[6]   TREATMENT OF CASTRATION-INDUCED OSTEOPOROSIS BY A CAPACITIVELY COUPLED ELECTRICAL SIGNAL IN RAT VERTEBRAE [J].
BRIGHTON, CT ;
LUESSENHOP, CP ;
POLLACK, SR ;
STEINBERG, DR ;
PETRIK, ME ;
KAPLAN, FS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1989, 71A (02) :228-236
[7]   PREVENTION AND TREATMENT OF SCIATIC DENERVATION DISUSE OSTEOPOROSIS IN THE RAT TIBIA WITH CAPACITIVELY COUPLED ELECTRICAL-STIMULATION [J].
BRIGHTON, CT ;
KATZ, MJ ;
GOLL, SR ;
NICHOLS, CE ;
POLLACK, SR .
BONE, 1985, 6 (02) :87-97
[8]   TREATMENT OF SCIATIC DENERVATION DISUSE OSTEOPOROSIS IN THE RAT TIBIA WITH CAPACITIVELY COUPLED ELECTRICAL-STIMULATION - DOSE-RESPONSE AND DUTY CYCLE [J].
BRIGHTON, CT ;
TADDUNI, GT ;
POLLACK, SR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1985, 67A (07) :1022-1028
[9]   RESPONSE OF CULTURED BONE-CELLS TO A CAPACITIVELY COUPLED ELECTRIC-FIELD - INHIBITION OF CAMP RESPONSE TO PARATHYROID-HORMONE [J].
BRIGHTON, CT ;
MCCLUSKEY, WP .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1988, 6 (04) :567-571
[10]  
BRIGHTON CT, 1992, CLIN ORTHOP RELAT R, P255