Comparison of patient doses in 256-slice CT and 16-slice CT scanners

被引:59
作者
Mori, S
Endo, M
Nishizawa, K
Murase, K
Fujiwara, H
Tanada, S
机构
[1] Natl Inst Radiol Sci, Dept Med Phys, Chiba 2638555, Japan
[2] Osaka Univ, Fac Med, Sch Allied Hlth Sci, Suita, Osaka 5650871, Japan
[3] Natl Inst Radiol Sci, Dept Med Imaging, Chiba 2638555, Japan
关键词
D O I
10.1259/bjr/39775216
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The 256-slice CT-scanner has been developed at the National Institute of Radiological Sciences. Nominal beam width was 128 mm in the longitudinal direction. When scanning continuously at the same position to obtain four-dimensional (4D) images, the effective dose is increased in proportion to the scan time. Our purpose in this work was to measure the dose for the 256-slice CT, to compare it with that of the 16-slice CT-scanner, and to make a preliminary assessment of dose for dynamic 3D imaging (volumetric cine imaging). Our group reported previously that the phantom length and integration range for dosimetry needed to be at least 300 mm to represent more than 90% of the line integral dose with the beam width between 20 mm and 138 mm. In order to obtain good estimates of the dose, we measured the line-integral dose over a 300 mm range in PMMA (polymethylmethacrylate) phantoms of 160 mm or 320 mm diameter and 300 mm length. Doses for both CT systems were compared for a clinical protocol. The results showed that the 256-slice CT generates a smaller dose than the 16-slice CT in all examinations. For volumetric cine imaging, we found an acceptable scan time would be 6s to 11s, depending on examinations, if dose must be limited to the same values as routine examinations with a conventional multidetector CT. Finally, we discussed the studies necessary to make full use of volumetric cine imaging.
引用
收藏
页码:56 / 61
页数:6
相关论文
共 13 条
[1]  
AOKI C, 2001, JPN J MED IMAGING, V20, P101
[2]   Development and performance evaluation of the first model of 4-D CT-scanner [J].
Endo, M ;
Mori, S ;
Tsunoo, T ;
Kandatsu, S ;
Tanada, S ;
Aradate, H ;
Saito, Y ;
Miyazaki, H ;
Satoh, K ;
Matsusita, S ;
Kusakabe, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (05) :1667-1671
[3]  
European Guidelines, 1997, 16262 EUR
[4]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[5]  
*INT COMM RAD PROT, 2000, ICRP PUBL ANN ICRP, V87
[6]  
*INT EL COMM, 1994, PUB IEC, P1223
[7]  
Kachelriess M, 1999, RADIOLOGY, V213P, P283
[8]   Enlarged longitudinal dose profiles in cone-beam CT and the need for modified dosimetry [J].
Mori, S ;
Endo, M ;
Nishizawa, K ;
Tsunoo, T ;
Aoyama, T ;
Fujiwara, H ;
Murase, K .
MEDICAL PHYSICS, 2005, 32 (04) :1061-1069
[9]   Clinical potentials of the prototype 256-detector row CT-scanner [J].
Mori, S ;
Endo, M ;
Obata, T ;
Murase, K ;
Fujiwara, H ;
Susumu, K ;
Tanada, S .
ACADEMIC RADIOLOGY, 2005, 12 (02) :148-154
[10]   Physical performance evaluation of a 256-slice CT-scanner for four-dimensional imaging [J].
Mori, S ;
Endo, M ;
Tsunoo, T ;
Kandatsu, S ;
Tanada, S ;
Aradate, H ;
Saito, Y ;
Miyazaki, H ;
Satoh, K ;
Matsushita, S ;
Kusakabe, M .
MEDICAL PHYSICS, 2004, 31 (06) :1348-1356