Dual-Modality Imaging of Cancer with SPECT/CT

被引:74
作者
Hasegawa, Bruce H. [1 ,2 ,3 ,4 ]
Wong, Kenneth H. [1 ,2 ,3 ,4 ]
Iwata, Koji [1 ,4 ]
Barber, William C. [1 ,4 ]
Hwang, Andrew B. [1 ,2 ,3 ]
Sakdinawat, Anne E. [1 ,4 ]
Ramaswamy, Mohan [4 ]
Price, David C. [4 ]
Hawkins, Randall A. [2 ,3 ,4 ]
机构
[1] Univ Calif San Francisco, Phys Res Lab, San Francisco, CA 94080 USA
[2] Univ Calif San Francisco, Bioengn Grad Grp, San Francisco, CA 94143 USA
[3] Univ Calif Berkeley, San Francisco, CA 94143 USA
[4] Univ Calif San Francisco, Dept Radiol, San Francisco, CA 94143 USA
关键词
radionuclide imaging; computed tomography; tumor imaging;
D O I
10.1177/153303460200100605
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Dual-modality imaging is an in vivo diagnostic technique that obtains structural and functional information directly from patient studies in a way that cannot be achieved with separate imaging systems alone. Dual-modality imaging systems are configured by combining computed tomography (CT) with radionuclide imaging (using positron emission tomography (PET) or single-photon emission computed tomography (SPECT)) on a single gantry which allows both functional and structural imaging to be performed during a single imaging session without having the patient leave the imaging system. A SPECT/CT system developed at UCSF is being used in a study to determine if dual-modality imaging offers advantages for assessment of patients with prostate cancer using In-111-ProstaScint (R), a radiolabeled antibody for the prostate-specific membrane antigen. In-111-ProstaScint (R) images are reconstructed using an iterative maximum-likelihood expectation-maximization (ML-EM) algorithm with correction for photon attenuation using a patient-specific map of attenuation coefficients derived from CT. The ML-EM algorithm accounts for the dual-photon nature of the In-111-labeled radionuclide, and incorporates correction for the geometric response of the radionuclide collimator. The radionuclide image then can be coregistered and overlaid in color on a grayscale CT image for improved localization of the functional information from SPECT. Radionuclide images obtained with SPECT/CT and reconstructed using ML-EM with correction for photon attenuation and collimator response improve image quality in comparison to conventional radionuclide images obtained with filtered backprojection reconstruction. These results illustrate the potential advantages of dual-modality imaging for improving the quality and the localization of radionuclide uptake for staging disease, planning treatment, and monitoring therapeutic response in patients with cancer.
引用
收藏
页码:449 / 458
页数:10
相关论文
共 69 条
[1]   QUANTITATIVE SPECT BY ATTENUATION CORRECTION OF THE PROJECTION SET USING TRANSMISSION DATA - EVALUATION OF A METHOD [J].
ALMQUIST, H ;
PALMER, J ;
LJUNGBERG, M ;
WOLLMER, P ;
STRAND, SE ;
JONSON, B .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1990, 16 (8-10) :587-594
[2]  
Bai C., 1997, J NUCL MED, V38, p214P
[3]  
BAILEY DL, 1987, J NUCL MED, V28, P844
[4]  
Beyer T, 2000, J NUCL MED, V41, P1369
[5]  
Beyer T., 1994, IEEE NSS MIC, V4, P1573
[6]  
Blankespoor S.C., 1996, THESIS U CALIFORNIA
[7]  
Blankespoor S. C., 1994, IEEE 1994 NUCL SCI S, V4, P1758
[8]   Attenuation correction of SPECT using X-ray CT on an emission-transmission CT system: Myocardial perfusion assessment [J].
Blankespoor, SC ;
Wu, X ;
Kalki, K ;
Brown, JK ;
Tang, HR ;
Cann, CE ;
Hasegawa, BH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (04) :2263-2274
[9]   Gamma camera-mounted anatomical X-ray tomography: technology, system characteristics and first images [J].
Bocher, M ;
Balan, A ;
Krausz, Y ;
Shrem, Y ;
Lonn, A ;
Wilk, M ;
Chisin, R .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2000, 27 (06) :619-627
[10]  
Brooks P. D., 1975, COMPUT TOMOG, V117, P561