A Y-channel design for improving zeta potential and surface conductivity measurements using the current monitoring method

被引:28
作者
Almutairi, Zeyad A. [1 ]
Glawdel, Tomasz [1 ]
Ren, Carolyn L. [1 ]
Johnson, David A. [1 ]
机构
[1] Univ Waterloo, Dept Mech & Mechatron Engn, Waterloo, ON N2L 3G1, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Zeta potential; Current monitoring technique; Y-channel design; Current slope method; PDMS microchannel; ELECTROOSMOTIC FLOW VELOCITY; CAPILLARY-ZONE-ELECTROPHORESIS; MICROCHANNELS; DEVICES; CONDUCTANCE; SYSTEMS; CHIP;
D O I
10.1007/s10404-008-0320-6
中图分类号
TB3 [工程材料学];
学科分类号
082905 [生物质能源与材料];
摘要
This study presents a new Y-channel design for measuring the zeta potential and surface conductance of a solid-liquid pairing using the current monitoring technique. The new design improves the throughput and reliability of the testing apparatus since the displacement between two solutions can be repeated many times without interfering with the experiments. It also increases the accuracy of the measurement by producing sharper start and end transitions for the current-time plot of the solution displacement process. In this design, efforts have been made to minimize the effects of electrolysis, Joule heating and undesired pressure driven flow on the measurements. An improvement on the current-time slope analysis is also presented. The Y-channel design was validated by comparing zeta potential measurements to published results. The zeta potential of several biological buffers relevant to the microfluidic community in plasma treated PDMS/PDMS and PDMS/Glass microchannels are presented. Preliminary studies of surface conductivity measurements using the Y-channel design were also conducted and are briefly discussed.
引用
收藏
页码:241 / 251
页数:11
相关论文
共 30 条
[1]
Determining ζ potential and surface conductance by monitoring the current in electro-osmotic flow [J].
Arulanandam, S ;
Li, DQ .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 2000, 225 (02) :421-428
[2]
Beckwith T.G., 1993, MECH MEASUREMENTS, V5 th
[4]
Electroosmotic flow in composite microchannels and implications in microcapillary electrophoresis systems [J].
Bianchi, F ;
Wagner, F ;
Hoffmann, P ;
Girault, HH .
ANALYTICAL CHEMISTRY, 2001, 73 (04) :829-836
[5]
A planar electroosmotic micropump [J].
Chen, CH ;
Santiago, JG .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2002, 11 (06) :672-683
[6]
CHOI ES, 2007, 7 INT C MIN CHEM BIO
[7]
An improved method of determining the ζ-potential and surface conductance [J].
Erickson, D ;
Li, DQ ;
Werner, C .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 2000, 232 (01) :186-197
[8]
Integrated capillary electrophoresis devices with an efficient postcolumn reactor in planar quartz and glass chips [J].
Fluri, K ;
Fitzpatrick, G ;
Chiem, N ;
Harrison, DJ .
ANALYTICAL CHEMISTRY, 1996, 68 (23) :4285-4290
[9]
Microfluidics - a review [J].
Gravesen, Peter ;
Branebjerg, Jens ;
Jensen, Ole Sondergard .
Journal of Micromechanics and Microengineering, 1993, 3 (04) :168-182
[10]
CAPILLARY ELECTROPHORESIS AND SAMPLE INJECTION SYSTEMS INTEGRATED ON A PLANAR GLASS CHIP [J].
HARRISON, DJ ;
MANZ, A ;
FAN, ZH ;
LUDI, H ;
WIDMER, HM .
ANALYTICAL CHEMISTRY, 1992, 64 (17) :1926-1932