Clinical experience with a dual-chamber implantable cardioverter defibrillator to treat atrial tachyarrhythmias

被引:65
作者
Gold, MR
Sulke, N
Schwartzman, DS
Mehra, R
Euler, DE
机构
[1] Univ Maryland, Baltimore, MD 21201 USA
[2] Eastbourne Hosp, Eastbourne, England
[3] Univ Pittsburgh, Atrial Arrhythmia Ctr, Pittsburgh, PA USA
[4] Medtronic Inc, Minneapolis, MN USA
关键词
implantable cardioverter defibrillator; atrial fibrillation; atrial tachycardia; pacing;
D O I
10.1046/j.1540-8167.2001.01247.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Dual-Chamber ICD and Atrial Tachyarrhythmia. Introduction: This study evaluated the safety and efficacy of a new dual-chamber implantable cardioverter defibrillator (ICD) to detect and treat atrial tachyarrhythmias in patients with drug-refractory atrial fibrillation (AF) and no indication for a ventricular ICD. Methods and Results: A dual-chamber ICD (Medtronic 7250 Jewel AF) was implanted in 144 of 146 patients. The device discriminates atrial tachycardia from AF based on cycle length and regularity, and uses atrial overdrive pacing as well as shocks to terminate tachyarrhythmia episodes. Patients were followed for an average of 12.6 +/-6.2 months. Use of antiarrhythmic drugs was 63% at baseline and did not change over time. Kaplan-Meier estimates of 12-month complication-free survival, device therapy survival, and patient survival were 85%, 91%, and 98%, respectively. Positive predictive accuracy of spontaneous atrial tachyarrhythmia detection was 99%, while atrial overdrive pacing and shocks terminated 40% and 87% of treated episodes, respectively. Median duration of successfully treated episodes was 8.9 minutes versus 144 minutes for the therapy failures. There was no reduction in the use of patient-activated shock therapy over time; at the 12-month follow-up evaluation, 94% of patients were in sinus rhythm. Ventricular tachyarrhythmias (67 episodes) were detected and appropriately treated in 7.6% of patients. Conclusion: This dual-chamber ICD appears to be safe and well tolerated in patients with drug-refractory symptomatic atrial tachyarrhythmias. The device, used in combination with drugs, effectively treats atrial tachyarrhythmias with pacing and/or shock therapies and decreases the median episode duration. In addition, the device protects from ventricular tachyarrhythmias in patients with AF and structural heart disease.
引用
收藏
页码:1247 / 1253
页数:7
相关论文
共 16 条
[1]   Efficacy of pacing therapies for treating atrial tachyarrhythmias in patients with ventricular arrhythmias receiving a dual-chamber implantable cardioverter defibrillator [J].
Adler, SW ;
Wolpert, C ;
Warman, EN ;
Musley, SK ;
Koehler, JL ;
Euler, DE .
CIRCULATION, 2001, 104 (08) :887-892
[2]   Electrophysiologic testing to identify patients with coronary artery disease who are at risk for sudden death [J].
Buxton, AE ;
Lee, KL ;
DiCarlo, L ;
Gold, MR ;
Greer, GS ;
Prystowsky, EN ;
O'Toole, MF ;
Tang, A ;
Fisher, JD ;
Coromilas, J ;
Talajic, M ;
Hafley, G .
NEW ENGLAND JOURNAL OF MEDICINE, 2000, 342 (26) :1937-1945
[3]   Initial clinical experience with ambulatory use of an implantable atrial defibrillator for conversion of atrial fibrillation [J].
Daoud, EG ;
Timmermans, C ;
Fellows, C ;
Hoyt, R ;
Lemery, R ;
Dawson, K ;
Ayers, GM .
CIRCULATION, 2000, 102 (12) :1407-1413
[4]   Dual chamber arrhythmia detection in the implantable cardioverter defibrillator [J].
Dijkman, B ;
Wellens, HJJ .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2000, 11 (10) :1105-1115
[5]   Atrial therapies reduce atrial arrhythmia burden in defibrillator patients [J].
Friedman, PA ;
Dijkman, B ;
Warman, EN ;
Xia, A ;
Mehra, R ;
Stanton, MS ;
Hammill, SC .
CIRCULATION, 2001, 104 (09) :1023-1028
[6]   Effects of High-Frequency Atrial Pacing in Atypical Atrial Flutter and Atrial Fibrillation [J].
Giorgberidze I. ;
Saksena S. ;
Mongeon L. ;
Mehra R. ;
Krol R.B. ;
Munsif A.N. ;
Mathew P. .
Journal of Interventional Cardiac Electrophysiology, 1997, 1 (2) :111-123
[7]   Catheter ablation for atrial fibrillation [J].
Jaïs, P ;
Shah, DC ;
Haïssaguerre, M ;
Hocini, M ;
Peng, JT ;
Clémenty, J .
ANNUAL REVIEW OF MEDICINE, 2000, 51 :431-441
[8]   Treatment strategies for atrial fibrillation [J].
Jung, F ;
DiMarco, JP .
AMERICAN JOURNAL OF MEDICINE, 1998, 104 (03) :272-286
[9]   The ablate and pace trial: A prospective study of catheter ablation of the AV conduction system and permanent pacemaker implantation for treatment of atrial fibrillation [J].
Kay, GN ;
Ellenbogen, KA ;
Giudici, M ;
Redfield, MM ;
Jenkins, LS ;
Mianulli, M ;
Wilkoff, B .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 1998, 2 (02) :121-135
[10]   STANDARDIZED REPORTING OF ICD PATIENT OUTCOME - THE REPORT OF A NORTH-AMERICAN SOCIETY-OF-PACING-AND-ELECTROPHYSIOLOGY POLICY CONFERENCE, FEBRUARY 9-10, 1993 [J].
KIM, SG ;
FOGOROS, RN ;
FURMAN, S ;
CONNOLLY, SJ ;
KUCK, KH ;
MOSS, AJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (07) :1358-1362