Effects of fluoride concentration and elastic tensile strain on the corrosion resistance of commercially pure titanium

被引:120
作者
Huang, HH [1 ]
机构
[1] Chung Shan Med & Dent Coll, Sch Dent, Inst Dent Mat, Taichung, Taiwan
关键词
fluoride; strain; corrosion resiatance; polarization resistance; titanium;
D O I
10.1016/S0142-9612(01)00079-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The corrosion resistance of commercially pure (CP) titanium in 1% NaCl divided by 0 similar to 1% NaF solution (pH = 6) under different elastic tensile strains was investigated by using electrochemical impedance spectroscopy (EIS) measurement technique. The polarization resistance (Rp), which is inversely proportional to the corrosion rate, of CP titanium in the test solution was obtained from the EIS data. Different elastic tensile strains, namely 0%, 1%, 2%, and 4%, were applied on the test specimens by using a tensile test machine during the corrosion tests. Corrosion morphology was characterized by using a scanning electron microscope. Surface chemical analyses were performed by using energy dispersive spectroscopy and X-ray photoelectron spectroscopy. The results showed that the NaF concentration and the elastic tensile strain had a statistically significant influence on the R-p (P <0.001). The R-p decreased on increasing the NaF concentration and the elastic tensile strain. When the NaF concentration was lower than 0.01 %, the R-p value ( > 3.4 x 10(5) Omega cm(2)) was mainly ascribed to the formation of a protective titanium dioxide (TiO2) on the metal surface, regardless of the elastic tensile strain applied. However, when the NaF concentration was higher than 0. 1%, the protectiveness of TiO2 was destroyed by fluoride ions, leading to severe corrosion of CP titanium. (C) 2001 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:59 / 63
页数:5
相关论文
共 11 条
[1]   INFLUENCE OF FLUORIDE ON TITANIUM IN AN ACIDIC ENVIRONMENT MEASURED BY POLARIZATION RESISTANCE TECHNIQUE [J].
BOERE, G .
JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (04) :283-288
[2]  
Jones D.A., 1996, PRINCIPLES PREVENTIO, Vsecond ed, P143
[3]   SEM OBSERVATIONS ON STRESS-CORROSION CRACKING OF COMMERCIALLY PURE TITANIUM IN A TOPICAL FLUORIDE SOLUTION [J].
KONONEN, MHO ;
LAVONIUS, ET ;
KIVILAHTI, JK .
DENTAL MATERIALS, 1995, 11 (04) :269-272
[4]   Effect of fluoride concentration and pH on corrosion behavior of titanium for dental use [J].
Nakagawa, M ;
Matsuya, S ;
Shiraishi, T ;
Ohta, M .
JOURNAL OF DENTAL RESEARCH, 1999, 78 (09) :1568-1572
[5]  
Probster L, 1992, INT J ORAL MAXILLOF, V7, P390
[6]  
SCHUTZ RW, 1987, METALS HDB, V13, P669
[7]  
Siirila H S, 1991, Int J Oral Maxillofac Implants, V6, P50
[8]   In vitro corrosion of titanium [J].
Strietzel, R ;
Hösch, A ;
Kalbfleisch, H ;
Buch, D .
BIOMATERIALS, 1998, 19 (16) :1495-1499
[9]   Corrosive properties of fluoride-containing odontologic gels against titanium [J].
ToumelinChemla, F ;
Rouelle, F ;
Burdairon, G .
JOURNAL OF DENTISTRY, 1996, 24 (1-2) :109-115