Peak distortion, data sampling errors and the integrator in the measurement of very narrow chromatographic peaks

被引:57
作者
Dyson, N [1 ]
机构
[1] The Croft, Dyson Instruments, Low Fell NE9 5YP, Tyne & Wear, England
关键词
peak shape; data sampling error; fast chromatography; electrochromatography; data processing;
D O I
10.1016/S0021-9673(99)00299-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
As more chromatographers consider using the techniques of fast chromatography or capillary electrochromatography, the instrumentation problems of generating and measuring these very narrow peaks is re-appraised. In particular the general view that 20-30 samples per peak is sufficient data to measure a peak is shown to be true under limited circumstances. A reworking of numerical integration theory and calculation of the errors of Newton Cotes rules when applied to chromatographic peaks, shows that asymmetry creates more subtle mischief: as many as 350 samples/peak may be needed to achieve 0.1% accuracy. Specialist, low time constant units are required to generate narrow peaks and a new breed of fast sampling data processor is required to measure them. As peaks narrow, it is increasingly important that data processors help analysts to identify data under-sampling by reporting peak asymmetry, actual sampling frequency las opposed to that initially programmed) and number of measured samples/peak as a combined validation diagnostic. Finally the article considers the lack of development of new deconvolution based data processors and points towards the lack of information inside flame ionisation and ultraviolet absorbance detectors. If new data processors were to become available for information rich detectors, their benefits: improved accuracy, precision and greater confidence in results, would have to be weighed against the costs of adopting them, re-working of analysis methods and retraining staff. Many production laboratories would find it uneconomical and would stay with the old methods. (C) 1999 Published by Elsevier Science B.V. All rights reserved.
引用
收藏
页码:321 / 340
页数:20
相关论文
共 41 条
[1]  
Annino R, 1996, HRC-J HIGH RES CHROM, V19, P285
[2]   INDUSTRIAL-GRADE FIELD-MOUNTABLE GAS-CHROMATOGRAPH FOR PROCESS MONITORING AND CONTROL [J].
ANNINO, R .
JOURNAL OF CHROMATOGRAPHY A, 1994, 678 (02) :279-288
[3]  
[Anonymous], NUMERICAL METHODS WO
[4]  
Beens J, 1998, HRC-J HIGH RES CHROM, V21, P47
[5]  
Burden RL, 1997, NUMERICAL ANAL
[6]   EFFECT OF PEAK SENSING AND RANDOM NOISE ON PRECISION AND ACCURACY OF STATISTICAL MOMENT ANALYSES FROM DIGITAL CHROMATOGRAPHIC DATA [J].
CHESLER, SN ;
CRAM, SP .
ANALYTICAL CHEMISTRY, 1971, 43 (14) :1922-&
[7]  
COE D, 1994, INT S CAP CHROM RIV
[8]   Advances in capillary electrochromatography: Rapid and high-efficiency separations of PAHs [J].
Dadoo, R ;
Zare, RN ;
Yan, C ;
Anex, DS .
ANALYTICAL CHEMISTRY, 1998, 70 (22) :4787-4792
[9]  
DAVIES ER, 1993, ELECT NOISE SIGNAL R
[10]   Maximum entropy chromatogram reconstruction [J].
Dimandja, JMD ;
Kaljurand, M ;
Phillips, JB ;
Valentín, J .
ANALYTICA CHIMICA ACTA, 1998, 371 (01) :1-8