Automated sequential injection fluorimetric determination and dissolution studies of Ergotamine Tartrate in pharmaceuticals

被引:22
作者
Legnerová, Z [1 ]
Sklenárová, H [1 ]
Solich, P [1 ]
机构
[1] Charles Univ, Fac Pharm, Dept Analyt Chem, Lab Flow Anal, Hradec Kralove 50005, Czech Republic
关键词
sequential injection analysis (SIA); dissolution test; ergotamine tartrate; fluorimetric detection;
D O I
10.1016/S0039-9140(02)00416-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A fully automated flow system for drug-dissolution studies based on the sequential injection analysis (SIA) was described and used for monitoring dissolution profiles of Ergotamine Tartrate (ET) in pharmaceutical formulations. 50 mul of dissolution medium was taken for each measurement at a flow rate of 40 mul s(-1) and detected by fluorescence detector using lambda(ex) = 236 nm (lambda(em) greater than or equal to 390 nm). The calibration curve was linear over the range 0.03-0.61 mg l(-1) of ET (sufficient for the dissolution tests). Equation of the calibration curve was calculated giving the following values: F= 117.7 c+0.80 (n = 6); r = 0.9998. Detection limit was 0.01 mg l(-1) of ET. The R.S.D. is less than 0.54 and 0.86% (n = 10) when determining 0.61 and 0.03 mg l(-1) of ET in standard solution, respectively. The dissolution test of Bellaspon tablets (0.3 mg of ET in 1 tablet) was programmed for 20 min, with a continuous sampling rate of 120 h under conditions required by BP 1993. (C) 2002 Published by Elsevier Science B.V.
引用
收藏
页码:1151 / 1155
页数:5
相关论文
共 14 条
[1]  
DECASTRO MD, 1995, J PHARM BIOMED ANAL, V8, P889
[2]   Continuous monitoring in drug dissolution testing using flow injection systems [J].
Fang, ZL ;
Fang, Q ;
Liu, XZ ;
Chen, HW ;
Liu, CL .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1999, 18 (04) :261-271
[3]   Trends of flow injection sample pretreatment approaching the new millennium [J].
Fang, ZL .
ANALYTICA CHIMICA ACTA, 1999, 400 :233-247
[4]   AUTOMATED FLOW-INJECTION TECHNIQUE FOR USE IN DISSOLUTION STUDIES OF SUSTAINED-RELEASE FORMULATIONS - APPLICATION TO IRON(II) FORMULATIONS [J].
GEORGIOU, CA ;
VALSAMI, GN ;
MACHERAS, PE ;
KOUPPARIS, MA .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1994, 12 (05) :635-641
[5]  
IZCZEFFY J, 1993, J PHARM BIOMED ANAL, V11, P191
[6]   AUTOMATED FLOW-INJECTION DETERMINATION OF SULFONAMIDES BY THE BRATTON-MARSHALL REACTION FOR CLINICAL ANALYSIS, ASSAYS AND DISSOLUTION STUDIES OF FORMULATIONS [J].
KOUPPARIS, MA ;
ANAGNOSTOPOULOU, PI .
ANALYTICA CHIMICA ACTA, 1988, 204 (1-2) :271-283
[7]   Sequential-injection system for drug-dissolution studies of ibuprofen tablets and sustained-release formulations [J].
Liu, XZ ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 1998, 358 (02) :103-110
[8]   Simultaneous monitoring of aspirin, phenacetin and caffeine in compound aspirin tablets using a sequential injection drug-dissolution testing system with partial least squares calibration [J].
Liu, XZ ;
Liu, SS ;
Wu, JF ;
Fang, ZL .
ANALYTICA CHIMICA ACTA, 1999, 392 (2-3) :273-281
[9]   Potentiometric determination of acetylsalicylic acid by sequential injection analysis (SIA) using a tubular salicylate-selective electrode [J].
Paseková, H ;
Sales, MG ;
Montenegro, MC ;
Araújo, AN ;
Polásek, M .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2001, 24 (5-6) :1027-1036
[10]  
POLASEK M, 1989, CESK FARM, V38, P1