Sequential-injection system for drug-dissolution studies of ibuprofen tablets and sustained-release formulations

被引:24
作者
Liu, XZ [1 ]
Fang, ZL [1 ]
机构
[1] Northeastern Univ, Dept Chem, Flow Inject Anal Res Ctr, Shenyang 110005, Peoples R China
关键词
sequential-injection analysis (SIA); dissolution test; ibuprofen; sustained-release formulation;
D O I
10.1016/S0003-2670(97)00602-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automated system for drug-dissolution studies based on the sequential-injection (SI) technique is described and evaluated for monitoring of dissolution profiles of ibuprofen tablets, sustained-release capsules and controlled-release matrix tablets. The system was composed of a computer-controlled syringe pump and a multi-position selection valve and coupled to a spectrophotometric detector. An 800 mu l test solution in dissolution vessel was aspirated sequentially into a PTFE holding coil via on-line filters. After delivering an aliquot (40 mu l) of each aspirated solution into the transferring line connected to the detector, the residual dissolution solution in the holding coil was fed back to the dissolution vessel to clear the filter and decrease sample consumption. The sampled aliquot was carried to the detector and monitored at 222 nm. Every hour, seven samples run in hexa-replicates were processed with a total of 42 measurements. The system was continuously run for 12 h with an overall relative standard deviation of 2.18% (n=72) at 50% final concentration level of the dissolution solution, which demonstrates the stability and robustness of the automated SI system. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:103 / 110
页数:8
相关论文
共 13 条
[1]  
Barnett NW, 1996, FRESEN J ANAL CHEM, V355, P591
[2]  
*CHEM IND PRESS CH, 1995, CHIN PHARM
[3]   AUTOMATION OF PHARMACEUTICAL DISSOLUTION TESTING BY FLOW-INJECTION ANALYSIS [J].
DECASTRO, MDL ;
VALCARCEL, M .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1990, 8 (04) :329-336
[4]   AUTOMATIC CONTINUOUS ONLINE MONITORING OF SALICYLIC-ACID AND ACETYLSALICYLIC-ACID IN PHARMACEUTICALS [J].
FERNANDEZ, JML ;
DECASTRO, MDL ;
VALCARCEL, M .
JOURNAL OF AUTOMATIC CHEMISTRY, 1990, 12 (06) :263-266
[5]   AUTOMATED FLOW-INJECTION TECHNIQUE FOR USE IN DISSOLUTION STUDIES OF SUSTAINED-RELEASE FORMULATIONS - APPLICATION TO IRON(II) FORMULATIONS [J].
GEORGIOU, CA ;
VALSAMI, GN ;
MACHERAS, PE ;
KOUPPARIS, MA .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1994, 12 (05) :635-641
[6]   MULTISTEP DETERMINATION OF ENZYME-ACTIVITY BY FLOW-INJECTION AND SEQUENTIAL INJECTION-ANALYSIS - ASSAY OF AMYLOGLUCOSIDASE [J].
HANSEN, EH ;
WILLUMSEN, B ;
WINTHER, SK ;
DRABOL, H .
TALANTA, 1994, 41 (11) :1881-1893
[7]   AUTOMATED FLOW-INJECTION DETERMINATION OF SULFONAMIDES BY THE BRATTON-MARSHALL REACTION FOR CLINICAL ANALYSIS, ASSAYS AND DISSOLUTION STUDIES OF FORMULATIONS [J].
KOUPPARIS, MA ;
ANAGNOSTOPOULOU, PI .
ANALYTICA CHIMICA ACTA, 1988, 204 (1-2) :271-283
[8]   THE AUTOMATION OF DISSOLUTION TESTING OF SOLID ORAL DOSAGE FORMS [J].
LAMPARTER, E ;
LUNKENHEIMER, C .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1992, 10 (10-12) :727-733
[9]   Sequential injection determination of D-glucose by chemiluminescence using an open tubular immobilised enzyme reactor [J].
Liu, XZ ;
Hansen, EH .
ANALYTICA CHIMICA ACTA, 1996, 326 (1-3) :1-12
[10]   SIMULTANEOUS MONITORING OF GLUCOSE, LACTIC-ACID AND PENICILLIN BY SEQUENTIAL INJECTION-ANALYSIS [J].
MIN, RW ;
NIELSEN, J ;
VILLADSEN, J .
ANALYTICA CHIMICA ACTA, 1995, 312 (02) :149-156