Effect of Certified Reference Material 470 (CRM 470) on national quality assurance programs for serum proteins in Europe

被引:38
作者
Johnson, AM
Whicher, JT
机构
[1] Univ N Carolina, Dept Pediat, Chapel Hill, NC USA
[2] Univ N Carolina, Dept Obstet & Gynecol, Chapel Hill, NC USA
[3] Rush House, Leeds, W Yorkshire, England
关键词
serum proteins; reference materials; quality control; standardization;
D O I
10.1515/CCLM.2001.177
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The introduction of the international reference material for serum proteins, CRM 470, has resulted in significant reduction of the among-laboratory variance for most proteins assayed in European national quality assurance programs. In general, the CVs have decreased by 5 to 65%. However, both among- and within-manufacturer variances in many cases remain unacceptably high. In addition, concentration-dependent differences in variance and bias are present for some proteins. Although some variance will persist, reducing variance and bias to levels required for the institution of universal reference ranges will necessitate more accurate transfer of values to calibrants and controls and improved calibration curve fitting by manufacturers, as well as better quality control within laboratories.
引用
收藏
页码:1123 / 1128
页数:6
相关论文
共 16 条