An observer study for direct comparison of clinical efficacy of electronic to film portal images

被引:16
作者
Yin, FF [1 ]
Rubin, P [1 ]
Schell, MC [1 ]
Wynn, R [1 ]
Raubertas, RF [1 ]
Uschold, G [1 ]
Sandhu, A [1 ]
Nelson, DF [1 ]
机构
[1] UNIV ROCHESTER, CTR CANC, DEPT BIOSTAT, ROCHESTER, NY 14642 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1996年 / 35卷 / 05期
关键词
electronic portal imaging device (EPID); portal film; portal verification; radiation therapy; observer study;
D O I
10.1016/0360-3016(96)00205-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To directly compare clinical efficacy of electronic to film portal images. Methods and Materials: An observer study was designed to compare clinical efficacy of electronic to film portal images acquired using a liquid matrix ion-chamber electronic portal imaging device and a conventional metal screen/film system, Both images were acquired simultaneously for each treatment port and the electronic portal images were printed on gray-level thermal paper, Four radiation oncologists served as observers and evaluated a total of 44 sets of images for four different treatment sites: lung, pelvis, brain, and head/neck. Each set of images included a simulation image, a double-exposure portal film, and video paper prints of electronic portal images, Eight to nine anatomical landmarks were selected from each treatment site, Each observer was asked to rate each landmark in terms of its clinical visibility and to rate the ease of making the pertinent verification decision in the corresponding electronic and film portal images with the aid of the simulation image. Results: Ratings for the visibility of landmarks and for the verification decision of treatment ports were similar for electronic and film images for most landmarks, However, vertebral bodies and several landmarks: in the pelvis such as the acetabulum and pubic symphysis were more visible in the portal film images than in the electronic portal images. Conclusion: The visibility. of landmarks in electronic portal images is comparable to that in film portal images. Verification of treatment ports based only on electronic portal images acquired using an electronic portal imaging device is generally achievable.
引用
收藏
页码:985 / 991
页数:7
相关论文
共 32 条
[1]  
[Anonymous], J AM STAT ASSOC
[2]  
[Anonymous], 1980, STAT METHODS CANC RE
[3]   DEMONSTRATION OF MEGAVOLTAGE AND DIAGNOSTIC-X-RAY IMAGING WITH HYDROGENATED AMORPHOUS-SILICON ARRAYS [J].
ANTONUK, LE ;
BOUDRY, J ;
HUANG, WD ;
MCSHAN, DL ;
MORTON, EJ ;
YORKSTON, J ;
LONGO, MJ ;
STREET, RA .
MEDICAL PHYSICS, 1992, 19 (06) :1455-1466
[4]   ONLINE REPOSITIONING DURING TREATMENT OF THE PROSTATE - A STUDY OF POTENTIAL LIMITS AND GAINS [J].
BALTER, JM ;
CHEN, GTY ;
PELIZZARI, CA ;
KRISHNASAMY, S ;
RUBIN, S ;
VIJAYAKUMAR, S .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 (01) :137-143
[5]   CORRELATION OF PROJECTION RADIOGRAPHS IN RADIATION-THERAPY USING OPEN CURVE SEGMENTS AND POINTS [J].
BALTER, JM ;
PELIZZARI, CA ;
CHEN, GTY .
MEDICAL PHYSICS, 1992, 19 (02) :329-334
[6]   A REVIEW OF ELECTRONIC PORTAL IMAGING DEVICES (EPIDS) [J].
BOYER, AL ;
ANTONUK, L ;
FENSTER, A ;
VANHERK, M ;
MEERTENS, H ;
MUNRO, P ;
REINSTEIN, LE ;
WONG, J .
MEDICAL PHYSICS, 1992, 19 (01) :1-16
[7]   WEEKLY LOCALIZATION FILMS AND DETECTION OF FIELD PLACEMENT ERRORS [J].
BYHARDT, RW ;
COX, JD ;
HORNBURG, A ;
LIERMANN, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1978, 4 (9-10) :881-887
[8]   QUALITY ASSURANCE USING PORTAL IMAGING - THE ACCURACY OF PATIENT POSITIONING IN IRRADIATION OF BREAST-CANCER [J].
CREUTZBERG, CL ;
ALTHOF, VGM ;
HUIZENGA, H ;
VISSER, AG ;
LEVENDAG, PC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (03) :529-539
[9]   INTERACTIVE USE OF ONLINE PORTAL IMAGING IN PELVIC RADIATION [J].
DENEVE, W ;
VANDENHEUVEL, F ;
COGHE, M ;
VERELLEN, D ;
DEBEUKELEER, M ;
ROELSTRAETE, A ;
DEROOVER, P ;
THON, L ;
STORME, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (03) :517-524
[10]   DAILY MONITORING AND CORRECTION OF RADIATION-FIELD PLACEMENT USING A VIDEO-BASED PORTAL IMAGING-SYSTEM - A PILOT-STUDY [J].
EZZ, A ;
MUNRO, P ;
PORTER, AT ;
BATTISTA, J ;
JAFFRAY, DA ;
FENSTER, A ;
OSBORNE, S .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 22 (01) :159-165