Design of near-optimal waveforms for chest and abdominal compression and decompression in CPR using computer-simulated evolution

被引:34
作者
Babbs, CF
机构
[1] Purdue Univ, Dept Basic Med Sci, W Lafayette, IN 47907 USA
[2] Indiana Univ, Sch Med, Indianapolis, IN 46204 USA
关键词
ACD-CPR; cardiopulmonary resuscitation (CPR); coronary perfusion pressure; device; IAC-CPR; mathematical model; vest CPR;
D O I
10.1016/j.resuscitation.2005.06.025
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To discover design principles underlying the optimal waveforms for external chest and abdominal compression and decompression during cardiac arrest and cardiopulmonary resuscitation (CPR). Method: A 14-compartment mathematical model of the human cardiopulmonary system is used to test successive generations of randomly mutated external compression waveforms during cardiac arrest and resuscitation. Mutated waveforms that produced superior mean perfusion pressure became parents for the next generation. Selection was based upon either systemic perfusion pressure (SPP=thoracic aortic minus right atrial pressure) or upon coronary perfusion pressure (CPP=thoracic aortic pressure minus myocardial wall pressure). After simulations of 64,414 individual CPR episodes, 40 highly evolved waveforms were characterized in terms of frequency, duty cycle, and phase. A simple, practical compression technique was then designed by combining evolved features with a constant rate of 80 min(-1) and duty cycle of 50%. Results: All ultimate surviving waveforms included reciprocal compression and decompression of the chest and the abdomen to the maximum allowable extent. The evolved waveforms produced 1.5-3 times the mean perfusion pressure of standard CPR and greater perfusion pressure than other forms of modified CPR reported heretofore, including active compression-decompression (ACD) + ITV and interposed abdominal compression (IAC)-CPR. When SPP was maximized by evolution, the chest compression/abdominal decompression phase was near 70% of cycle time. When CPP was maximized, the abdominal compression/chest decompression phase was near 30% of cycle time. Near-maximal SPP/CPP of 60/21 mmHg (forward flow 3.8 L/min) occurred at a compromise compression frequency of 80 min(-1) and duty cycle for chest compression of 50%.
引用
收藏
页码:277 / 293
页数:17
相关论文
共 68 条
[21]   CARDIOPULMONARY-RESUSCITATION RESEARCH MODELS - MOTHERBOARDS, MAMMALS, AND MAN - COMMENT [J].
CHRISTENSON, JM .
ACADEMIC EMERGENCY MEDICINE, 1995, 2 (08) :669-671
[22]   HYSTERETIC BEHAVIOR OF SOFT LIVING ANIMAL TISSUE [J].
CHU, BM ;
FRASHER, WG ;
WAYLAND, H .
ANNALS OF BIOMEDICAL ENGINEERING, 1972, 1 (02) :182-203
[23]  
DELGUERCIO LR, 1965, CIRCULATION, V31, P171
[24]   INHIBITION OF CORONARY BLOOD-FLOW BY A VASCULAR WATERFALL MECHANISM [J].
DOWNEY, JM ;
KIRK, ES .
CIRCULATION RESEARCH, 1975, 36 (06) :753-760
[25]   A two-layered mechanical model of the rat esophagus. Experiment and theory [J].
Fan, Yanhua ;
Gregersen, Hans ;
Kassab, Ghassan S. .
BIOMEDICAL ENGINEERING ONLINE, 2004, 3 (1)
[26]   TONE-DEPENDENT WATERFALL BEHAVIOR DURING VENOUS-PRESSURE ELEVATION IN ISOLATED CANINE HEARTS [J].
FARHI, ER ;
KLOCKE, FJ ;
MATES, RE ;
KUMAR, K ;
JUDD, RM ;
CANTY, JM ;
SATOH, S ;
SEKOVSKI, B .
CIRCULATION RESEARCH, 1991, 68 (02) :392-401
[27]  
Finlay J. B, 1973, PERSPECTIVES BIOMEDI, P165
[28]   CARDIAC-OUTPUT DURING CARDIOPULMONARY RESUSCITATION AT VARIOUS COMPRESSION RATES AND DURATIONS [J].
FITZGERALD, KR ;
BABBS, CF ;
FRISSORA, HA ;
DAVIS, RW ;
SILVER, DI .
AMERICAN JOURNAL OF PHYSIOLOGY, 1981, 241 (03) :H442-H448
[29]  
Fogel Q., 1966, ARTIF INTELL, P170
[30]  
FUNG YC, 1981, BIOMECHANICS MECH PR, P23