Early migration predicts late aseptic failure of hip sockets

被引:133
作者
Krismer, M [1 ]
Stockl, B [1 ]
Fischer, M [1 ]
Bauer, R [1 ]
Mayrhofer, P [1 ]
Ogon, M [1 ]
机构
[1] UNIV INNSBRUCK, ORTHOPAED HOSP, A-6020 INNSBRUCK, AUSTRIA
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1996年 / 78B卷 / 03期
关键词
D O I
10.1302/0301-620X.78B3.0780422
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report a prospective, stratified study of 60 PCA-cups and 60 RM-polyethylene cups which have been followed for a median time of 90 months, with annual radiography. The radiological migration of cups was measured by the computer-assisted EBRA method. A number of threshold migration rates from 1 nun in the first year to 1 mm in five years have been assessed and related to clinically determined revision rates. A total of 28 cups showed a total migration of 1 mm or more within the first two years; 13 of these cups have required revision and been exchanged. The survival curves of cups which had previously shown early migration were considerably different from those without early migration, For cups with a migration of less than 1 mm within the first two years the mean survival at 96 months was 0.96 +/- 0.02; for migrating cups, it was 0.63 +/- 0.11 (log-rank test, p = 0.0001; chi-square value = 39.4). Early migration is a good predictor for late loosening of hip sockets.
引用
收藏
页码:422 / 426
页数:5
相关论文
共 16 条
[11]   FIXATION AND LOOSENING OF HIP PROSTHESES - A REVIEW [J].
MJOBERG, B .
ACTA ORTHOPAEDICA SCANDINAVICA, 1991, 62 (05) :500-508
[12]  
ONSTEN I, 1995, CLIN ORTHOP RELAT R, P103
[13]   MIGRATION OF ACETABULAR COMPONENTS, INSERTED WITH AND WITHOUT CEMENT, IN ONE-STAGE BILATERAL HIP-ARTHROPLASTY - A CONTROLLED, RANDOMIZED STUDY USING ROENTGENSTEREOPHOTOGRAMMETRIC ANALYSIS [J].
ONSTEN, I ;
CARLSSON, AS ;
OHLIN, A ;
NILSSON, JA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1994, 76A (02) :185-194
[14]  
RUSSE W, 1988, RONTGENPHOTOGRAMMETR
[15]   PRIMARY MIGRATION OF FULLY-THREADED ACETABULAR PROSTHESES - A ROENTGEN STEREOPHOTOGRAMMETRIC ANALYSIS [J].
SNORRASON, F ;
KARRHOLM, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (04) :647-652
[16]   MIGRATION OF HYDROXYAPATITE-COATED FEMORAL PROSTHESES - A ROENTGEN STEREOPHOTOGRAMMETRIC STUDY [J].
SOBALLE, K ;
TOKSVIGLARSEN, S ;
GELINECK, J ;
FRUENSGAARD, S ;
HANSEN, ES ;
RYD, L ;
LUCHT, U ;
BUNGER, C .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (05) :681-687