Artifact reduction strategies for prosthetic heart valve CT imaging

被引:20
作者
Habets, Jesse [1 ]
Symersky, Petr [2 ]
Leiner, Tim [1 ]
de Mol, Bas A. J. M. [2 ]
Mali, Willem P. Th M. [1 ]
Budde, Ricardo P. J. [1 ,3 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiol, NL-3508 GA Utrecht, Netherlands
[2] Univ Amsterdam, Acad Med Ctr, Dept Cardiothorac Surg, NL-1105 AZ Amsterdam, Netherlands
[3] Gelre Hosp, Dept Radiol, Apeldoorn, Netherlands
关键词
Computed tomography; Prosthetic heart valve; Metal artifact reduction; Iterative reconstruction; Tube voltage; COMPUTED-TOMOGRAPHY; ECHOCARDIOGRAPHY;
D O I
10.1007/s10554-012-0041-5
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
Multislice CT evaluation of prosthetic heart valves (PHV) is limited by PHV-related artifacts. We assessed the influence of different kV settings, a metal artifact reduction filter (MARF) and an iterative reconstruction algorithm (IR) on PHV-induced artifacts in an in vitro model. A Medtronic-Hall tilting disc and St Jude bileafet PHV were imaged using a 64-slice scanner with 100 kV/165 mAs, 120 kV/100 mAs, 140 kV/67 mAs at an equal CTDIvol. Images were reconstructed with (1) filtered back projection (FBP), (2) IR, (3) MARF and (4) MARF and IR. Hypo- and hyperdense artifacts volumes (mean mm(3) +/- SD) were quantified with 2 thresholds (<=-50 and >= 175 Hounsfield Units). Image noise was measured and the presence of secondary artifacts was scored by 2 observers independently. Mean hypodense artifacts for the Medtronic-Hall/St Jude valve (FBP) were 966 +/- 23/1,738 +/- 21 at 100 kV, 610 +/- 13/991 +/- 12 at 120 kV, and 420 +/- 9/634 +/- 9 at 140 kV. Compared to FBP, hypodense artifact reductions for IR were 9/8 %, 10/7 % and 12/6 % respectively, for MARF 92 %/84 %, 89/81 % and 86/77 % respectively; for MARF + IR 94/85 %, 92/82 %, and 90/79 % respectively. Mean hyperdense artifacts for the Medtronic-Hall/St Jude valve were 5,530 +/- 48/6,940 +/- 70 at 100 kV, 5,120 +/- 42/6,250 +/- 53 at 120 kV, and 5,011 +/- 52/6,000 +/- 0 at 140 kV. Reductions for IR were 2/2 %, 2/3 % and 3/4 % respectively, for MARF were 9/30 %, 0/25 %, 5/22 % respectively, MARF + IR 12/32 %, 4/27 % and 7/25 % respectively. Secondary artifacts were found in all MARF images. Image noise was reduced in the IR images. In vitro PHV-related artifacts can be reduced by increasing kV despite maintaining identical CTDIvol. Although MARF is more effective than IR, it induces secondary artifacts.
引用
收藏
页码:2099 / 2108
页数:10
相关论文
共 21 条
[1]
Metal artifact reduction in CT using tissue-class modeling and adaptive prefiltering [J].
Bal, Matthieu ;
Spies, Lothar .
MEDICAL PHYSICS, 2006, 33 (08) :2852-2859
[2]
Artifacts in CT: Recognition and avoidance [J].
Barrett, JF ;
Keat, N .
RADIOGRAPHICS, 2004, 24 (06) :1679-1691
[3]
Evaluation of Two Iterative Techniques for Reducing Metal Artifacts in Computed Tomography [J].
Boas, F. Edward ;
Fleischmann, Dominik .
RADIOLOGY, 2011, 259 (03) :894-902
[4]
Evaluation of Anatomic Valve Opening and Leaflet Morphology in Aortic Valve Bioprosthesis by Using Multidetector CT: Comparison with Transthoracic Echocardiography [J].
Chenot, Fabien ;
Montant, Patrick ;
Goffinet, Celine ;
Pasquet, Agnes ;
Vancraeynest, David ;
Coche, Emmanuel ;
Vanoverschelde, Jean-Louis ;
Gerber, Bernhard L. .
RADIOLOGY, 2010, 255 (02) :377-385
[5]
Multislice Computed Tomography in Infective Endocarditis Comparison With Transesophageal Echocardiography and Intraoperative Findings [J].
Feuchtner, Gudrun M. ;
Stolzmann, Paul ;
Dichtl, Wolfgang ;
Schertler, Thomas ;
Bonatti, Johannes ;
Scheffel, Hans ;
Mueller, Silvana ;
Plass, Andre ;
Mueller, Ludwig ;
Bartel, Thomas ;
Wolf, Florian ;
Alkadhi, Hatem .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 53 (05) :436-444
[6]
Computed tomography-old ideas and new technology [J].
Fleischmann, Dominik ;
Boas, F. Edward .
EUROPEAN RADIOLOGY, 2011, 21 (03) :510-517
[7]
Reoperation for prosthetic aortic valve obstruction in the era of echocardiography: Trends in diagnostic testing and comparison with surgical findings [J].
Girard, SE ;
Miller, FA ;
Orszulak, TA ;
Mullany, CJ ;
Montgomery, S ;
Edwards, WD ;
Tazelaar, HD ;
Malouf, JF ;
Tajik, AJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (02) :579-584
[8]
A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves [J].
Habets, Jesse ;
Symersky, Petr ;
de Mol, Bas A. J. M. ;
Mali, Willem P. Th. M. ;
Leiner, Tim ;
Budde, Ricardo P. J. .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2012, 28 (06) :1567-1575
[9]
Diagnostic evaluation of left-sided prosthetic heart valve dysfunction [J].
Habets, Jesse ;
Budde, Ricardo P. ;
Symersky, Petr ;
van den Brink, Renee B. ;
de Mol, Bas A. ;
Mali, Willem P. ;
van Herwerden, Lex A. ;
Chamuleau, Steven A. .
NATURE REVIEWS CARDIOLOGY, 2011, 8 (08) :466-478
[10]
Prosthetic heart valve assessment with multidetector-row CT: imaging characteristics of 91 valves in 83 patients [J].
Habets, Jesse ;
Symersky, Petr ;
van Herwerden, Lex A. ;
de Mol, Bas A. J. M. ;
Spijkerboer, Anje M. ;
Mali, Willem P. Th. M. ;
Budde, Ricardo P. J. .
EUROPEAN RADIOLOGY, 2011, 21 (07) :1390-1396