Ratiometric fiber-optic oxygen sensors based on sol-gel matrix doped with metalloporphyrin and 7-amino-4-trifluoromethyl coumarin

被引:57
作者
Chu, Chen-Shane [1 ]
Lo, Yu-Lung [1 ]
机构
[1] Natl Cheng Kung Univ, Dept Mech Engn, Tainan 70101, Taiwan
关键词
Ratiometric; Fiber-optic oxygen sensor; Sol-gel; Metalloporphyrin;
D O I
10.1016/j.snb.2008.06.022
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper develops highly sensitive ratiometric fiber-optic oxygen sensors incorporating a sol-gel matrix doped with platinum or palladium tetrakis pentafluorophenyl porphine (PtTFPP or PdTFPP) metalloporphyrins as the oxygen-sensitive material and 7-amino-4-trifluoromethyl coumarin (AFC) as the reference dye. Using an LED with a central wavelength of 400nm as an excitation source, it is shown that the emission wavelengths of the oxygen-sensitive dye and the reference dye have no spectral overlap and therefore permit the oxygen concentration to be measured using a rationnetric-based method. The sensitivities of the PtTFPP-doped and PdTFPP-doped oxygen sensors are evaluated in terms of the ratio I-N2/I-O2 where I-N2 and I-O2 correspond to the detected luminescence intensities in pure nitrogen and pure oxygen, respectively. The experimental results reveal that the PtTFPP and PdTFPP sensors both yield linear Stern-Volmer plots and have sensitivities of around 33 and 74, respectively. In addition, the response time of the PtTFPP-doped sensor is found to be 2.8 s when switching from a fully deoxygenated environment (i.e. 100% nitrogen) to a fully oxygenated environment, and 5.8s when switching in the reverse direction. The corresponding response times of the PdTFPP-doped sensor are 7.9 and 58.9 s, respectively. The ratiometric sensing approach presented in this study not only yields high sensitivity and reversibility, but also has the advantage of suppressing the effects of spurious fluctuations in the intensity of the excitation source and optical transmission properties of the optic fiber. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:711 / 717
页数:7
相关论文
共 31 条
[1]  
ALLEN G, 1992, Patent No. 5094959
[2]   Probes and polymers for optical sensing of oxygen [J].
Amao, Y .
MICROCHIMICA ACTA, 2003, 143 (01) :1-12
[3]   Platinum tetrakis(pentafluorophenyl)porphyrin immobilized in polytrifluoroethylmethacrylate film as a photostable optical oxygen detection material [J].
Amao, Y ;
Miyashita, T ;
Okura, I .
JOURNAL OF FLUORINE CHEMISTRY, 2001, 107 (01) :101-106
[4]  
[Anonymous], 1999, PRINCIPLES FLUORESCE
[5]   Temperature-sensitive europium(III) probes and their use for simultaneous luminescent sensing of temperature and oxygen [J].
Borisov, Sergey M. ;
Wolfbeis, Otto S. .
ANALYTICAL CHEMISTRY, 2006, 78 (14) :5094-5101
[6]   Poly(decyl methacrylate)-based fluorescent PEBBLE swarm nanosensors for measuring dissolved oxygen in biosamples [J].
Cao, YF ;
Koo, YEL ;
Kopelman, R .
ANALYST, 2004, 129 (08) :745-750
[7]   PHOTOPHYSICS AND PHOTOCHEMISTRY OF OXYGEN SENSORS BASED ON LUMINESCENT TRANSITION-METAL COMPLEXES [J].
CARRAWAY, ER ;
DEMAS, JN ;
DEGRAFF, BA ;
BACON, JR .
ANALYTICAL CHEMISTRY, 1991, 63 (04) :337-342
[8]   High-performance fiber-optic oxygen sensors based on fluorinated xerogels doped with Pt(II) complexes [J].
Chu, Chen-Shane ;
Lo, Yu-Lung .
SENSORS AND ACTUATORS B-CHEMICAL, 2007, 124 (02) :376-382
[9]  
CLARK LC, 1956, T AM SOC ART INT ORG, V2, P41
[10]   Oxygen sensors based on luminescence quenching [J].
Demas, JN ;
DeGraff, BA ;
Coleman, PB .
ANALYTICAL CHEMISTRY, 1999, 71 (23) :793A-800A