Detecting endoleaks after endovascular AAA repair with a minimally invasive, implantable, telemetric pressure sensor: an in vitro study

被引:20
作者
Springer, Fabian
Schlierf, Roland
Pfeffer, Joachim-Georg
Mahnken, Andreas H.
Schnakenberg, Uwe
Schmitz-Rode, Thomas
机构
[1] Rhein Westfal TH Aachen Univ, Helmholtz Inst Biomed Engn Appl Med Engn, D-52074 Aachen, Germany
[2] Rhein Westfal TH Aachen Univ, Inst Mat Elect Engn 1, D-52074 Aachen, Germany
[3] Rhein Westfal TH Aachen Univ, Univ Hosp, Dept Diagnost Radiol, D-52074 Aachen, Germany
关键词
aortic aneurysm; endovascular repair; endoleak; pressure sensor;
D O I
10.1007/s00330-007-0583-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A feasibility study on a completely digital telemetric pressure sensor (TPS) to detect endoleaks was performed in an in vitro model of an abdominal aortic aneurysm (AAA). An endovascular-stented AAA silicone model with different types (I-III) and sizes (3-11 French) of endoleaks was created and pulsatile pressure was applied with physiological flow and pressure rates [mean intraaortic pressure (IAP): 95-130 mmHg] and different degrees of thrombosis of the aneurysm sac. Aneurysm sac pressure (ASP) was measured with the TPS and with wired pressure sensors (WPS) as a reference. Statistical analysis included paired t-test, Pearson's correlation analysis and Bland-Altman plots. After opening an endoleak, the mean ASP increased significantly (P < 0.0001) from 15 to almost 95% of the mean IAP depending on endoleak type and size. ASP could be measured accurately with the TPS and the WPS. The telemetric and wired ASP increase showed a high Pearson's correlation coefficient (r) for a non-thrombosed (r = 0.97) and a thrombosed (r = 0.96) aneurysm sac. In an in vitro silicone model, the newly designed telemetric pressure sensor was able to detect the occurrence of an endoleak in a non-invasive way and might be a valuable device for follow-up of endovascular AAA repair.
引用
收藏
页码:2589 / 2597
页数:9
相关论文
共 33 条
[1]  
ALLEN MG, 2005, 13 INT C SOL STAT SE, P275
[2]   Aneurysm sac pressure measurements after endovascular repair of abdominal aortic aneurysms [J].
Baum, RA ;
Carpenter, RP ;
Cope, C ;
Golden, MA ;
Velazquez, OC ;
Neschis, DG ;
Mitchell, ME ;
Barker, CF ;
Fairman, RM .
JOURNAL OF VASCULAR SURGERY, 2001, 33 (01) :32-40
[3]  
Chong CK, 2003, J ENDOVASC THER, V10, P424, DOI 10.1583/1545-1550(2003)010<0424:MEACRF>2.0.CO
[4]  
2
[5]   Intra-aneurysm sac pressure measurements after endovascular aneurysm repair: Differences between shrinking, unchanged, and expanding aneurysms with and without endoleaks [J].
Dias, NV ;
Ivancev, K ;
Malina, M ;
Resch, T ;
Lindblad, B ;
Sonesson, B .
JOURNAL OF VASCULAR SURGERY, 2004, 39 (06) :1229-1235
[6]  
Dubenec SR, 2003, J CARDIOVASC SURG, V44, P553
[7]   Abdominal aortic aneurysm sac shrinkage after endovascular aneurysm repair: Correlation with chronic sac pressure measurement [J].
Ellozy, SH ;
Carroccio, A ;
Lookstein, RA ;
Jacobs, TS ;
Addis, MD ;
Teodorescu, VJ ;
Marin, ML .
JOURNAL OF VASCULAR SURGERY, 2006, 43 (01) :2-6
[8]   First experience in human beings with a permanently implantable intrasac pressure transducer for monitoring endovascular repair of abdominal aortic aneurysms [J].
Ellozy, SH ;
Carroccio, A ;
Lookstein, RA ;
Minor, ME ;
Sheahan, CM ;
Juta, J ;
Cha, A ;
Valenzuela, R ;
Addis, MD ;
Jacobs, MS ;
Teodorescu, VJ ;
Marin, ML .
JOURNAL OF VASCULAR SURGERY, 2004, 40 (03) :405-411
[9]   Endotension as a result of pressure transmission through the graft following endovascular aneurysm repair - An in vitro study [J].
Gawenda, M ;
Jaschke, G ;
Winter, S ;
Wassmer, G ;
Brunkwall, J .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2003, 26 (05) :501-505
[10]  
Gilling-Smith G, 1999, J ENDOVASC SURG, V6, P305, DOI 10.1583/1074-6218(1999)006<0305:EAEARD>2.0.CO