Finite amplitude distortion and its relationship to linear derating formulae for diagnostic ultrasound systems

被引:46
作者
Christopher, T
Carstensen, EL
机构
[1] UNIV ROCHESTER,ROCHESTER CTR BIOMED ULTRASOUND,ROCHESTER,NY 14627
[2] UNIV ROCHESTER,DEPT ELECT ENGN,ROCHESTER,NY 14627
关键词
biomedical ultrasound; nonlinear propagation; finite amplitude effects; diagnostic ultrasound; derating; ultrasonic heating;
D O I
10.1016/S0301-5629(96)00099-3
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Formulation of indices that can be used as predictors of biological effects of ultrasound involves a process called derating, in which measurements of the sound field made in water are extrapolated to estimates of the magnitude of the sound fields in the tissues of the body, All indices that have been formulated up to the present time assume that the propagation of ultrasound is linear, In fact, under most exposure conditions for which biological effects may be a concern, sound propagation is highly nonlinear, A nonlinear propagation model has been used in this study to evaluate the nature of the effects that occur under realistic exposure conditions encountered in diagnostic procedures, Because of the way that the thermal index is defined, it turns out that ignoring nonlinear propagation leads to underestimates of tissue temperature increments that typically are less than 40%, As currently implemented, the mechanical index may be underestimated by more than a factor of two because it ignores the saturation of the sound fields that result from nonlinear propagation, For large propagation distances in soft tissues (e.g., 10 cm at 3 MHz in liver); however, it is physically difficult to exceed tissue pressures corresponding to MI > 2 because of these same saturation phenomena. Copyright (C) 1996 World Federation for Ultrasound in Medicine & Biology
引用
收藏
页码:1103 / 1116
页数:14
相关论文
共 36 条
[21]   LOSSES IN TISSUE ASSOCIATED WITH FINITE-AMPLITUDE ULTRASOUND TRANSMISSION [J].
FRY, FJ ;
DINES, KA ;
REILLY, CR ;
GOSS, SA .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1989, 15 (05) :481-497
[22]   COMPILATION OF EMPIRICAL ULTRASONIC PROPERTIES OF MAMMALIAN-TISSUES .2. [J].
GOSS, SA ;
JOHNSTON, RL ;
DUNN, F .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1980, 68 (01) :93-108
[23]   DEMONSTRATION OF ENHANCED TEMPERATURE ELEVATION DUE TO NONLINEAR PROPAGATION OF FOCUSED ULTRASOUND IN DOGS THIGH INVIVO [J].
HYNYNEN, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1987, 13 (02) :85-91
[24]  
KAMAKURA T, 1992, J ACOUST SOC AM, V90, P3144
[25]   DETERMINATION OF THE NONLINEARITY PARAMETER B/A OF BIOLOGICAL MEDIA [J].
LAW, WK ;
FRIZZELL, LA ;
DUNN, F .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1985, 11 (02) :307-318
[26]   ABSORPTION AND ATTENUATION IN SOFT-TISSUES .2. EXPERIMENTAL RESULTS [J].
LYONS, ME ;
PARKER, KJ .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1988, 35 (04) :511-521
[27]  
National Council on Radiation Protection and Measurement (NCRP), 1992, 113 NCRP
[28]   SOLUTIONS OF THE BIO-HEAT TRANSFER EQUATION [J].
NYBORG, WL .
PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (07) :785-792
[29]   ULTRASONIC-ATTENUATION AND ABSORPTION IN LIVER-TISSUE [J].
PARKER, KJ .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1983, 9 (04) :363-369
[30]  
SCHWAN HP, 1954, ELECTRONICS, P35