In-vitro characterisation of the nebulised dose during non-invasive ventilation

被引:74
作者
Abdelrahim, Mohamed E. [2 ,3 ]
Plant, Paul [4 ]
Chrystyn, Henry [1 ]
机构
[1] Univ Huddersfield, Sch Appl Sci, Div Pharm & Pharmaceut Sci, Huddersfield HD1 3DH, W Yorkshire, England
[2] Univ Bradford, Inst Pharmaceut Innovat, Bradford BD7 1DP, W Yorkshire, England
[3] Univ Bani Suef, Fac Pharm, Bani Suwayf, Egypt
[4] St James Univ Hosp, Leeds, W Yorkshire, England
关键词
aerodynamic characteristics; nebuliser; Next Generation Impactor; non-invasive ventilation; terbutaline; POSITIVE-PRESSURE VENTILATION; BRONCHODILATOR DELIVERY; AEROSOL; SALBUTAMOL; IMPACTOR; INHALER;
D O I
10.1111/j.2042-7158.2010.01134.x
中图分类号
R9 [药学];
学科分类号
100702 [药剂学];
摘要
Objectives Non-invasive ventilation (NIV) with nebulised bronchodilators helps some patients to maintain effective ventilation. However, the position of the nebuliser in the ventilation circuit may affect lung delivery. Methods We placed the nebuliser proximal (A) and distal (B) to a breathing simulator in a standard NIV circuit with inspiratory (I) and expiratory (E) pressures of 20 and 5 cm H2O, 1 : 3 I : E ratio, 15 breaths/min and a tidal volume of 500 ml. Five milligrams of terbutaline solution was nebulised using an Aeroneb Pro (AERO) and a Sidestream (SIDE) nebuliser. The fate of the nebulised dose was determined and the aerodynamic droplet characteristics were measured using a cooled Next Generation Impactor. Key findings More terbutaline was entrained on the inhalation filter in position A than in position B (P < 0.001) for both nebulisers. These amounts were greater (P < 0.001) for AERO than SIDE due to a smaller (P < 0.001) residual volume. The mean (SD) fine particle doses for AEROA, AEROB, SIDEA and SIDEB were 1.31 (0.2), 1.13 (0.14), 0.56 (0.03) and 0.39 (0.13) mg. These amounts from AEROA were significantly greater (P < 0.001) than those of the other three methods. Conclusions The results highlight the differences between nebulisers and the influence on the placement of the nebuliser in the NIV circuit.
引用
收藏
页码:966 / 972
页数:7
相关论文
共 25 条
[1]
Aerodynamic Characteristics of Nebulized Terbutaline Sulphate Using the Next Generation Impactor (NGI) and CEN Method [J].
Abdelrahim, Mohamed E. ;
Chrystyn, Henry .
JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2009, 22 (01) :19-28
[2]
BERG E, 2008, PHARMEUROPA SCI NOTE, V1, P21
[3]
Determination of nebulizer droplet size distribution: A method based on impactor refrigeration [J].
Berg, Elna ;
Svensson, Jan Olof ;
Asking, Lars .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2007, 20 (02) :97-104
[4]
Boe J, 2001, EUR RESPIR J, V18, P228
[5]
Branconnier Matthew P, 2005, Respir Care, V50, P1649
[6]
Enhanced delivery of nebulised salbutamol during non-invasive ventilation [J].
Calvert, Lori D. ;
Jackson, Judith M. ;
White, Judy A. ;
Barry, Peter W. ;
Kinnear, William J. ;
O'Callaghan, Christopher .
JOURNAL OF PHARMACY AND PHARMACOLOGY, 2006, 58 (11) :1553-1557
[7]
In vitro evaluation of aerosol bronchodilator delivery during noninvasive positive pressure ventilation:: Effect of ventilator settings and nebulizer position [J].
Chatmongkolchart, S ;
Schettino, GPP ;
Dillman, C ;
Kacmarek, RM ;
Hess, DR .
CRITICAL CARE MEDICINE, 2002, 30 (11) :2515-2519
[8]
NASAL POSITIVE PRESSURE VENTILATION IN PATIENTS WITH ACUTE RESPIRATORY-FAILURE - DIFFICULT AND TIME-CONSUMING PROCEDURE FOR NURSES [J].
CHEVROLET, JC ;
JOLLIET, P ;
ABAJO, B ;
TOUSSI, A ;
LOUIS, M .
CHEST, 1991, 100 (03) :775-782
[9]
WASTAGE OF DRUG FROM NEBULIZERS - A REVIEW [J].
CLAY, MM ;
CLARKE, SW .
JOURNAL OF THE ROYAL SOCIETY OF MEDICINE, 1987, 80 (01) :38-39
[10]
*COM EUR NORM, 2001, 135441 PREN, P33