Effect of precision error on T-scores and the diagnostic classification of bone status

被引:23
作者
Kiebzak, Gary M.
Faulkner, Kenneth G.
Wacker, Wynn
Hamdy, Ronald
Seier, Edith
Watts, Nelson B.
机构
[1] St Lukes Belmont Ctr, Ctr Orthopaed Res & Educ, Houston, TX 77025 USA
[2] GE Healthcare, Madison, WI USA
[3] Synarc, San Francisco, CA USA
[4] E Tennessee State Univ, Coll Med, Osteoporosis Ctr, Johnson City, TN 37614 USA
[5] VAMC, Johnson City, TN 37614 USA
[6] E Tennessee State Univ, Dept Math, Johnson City, TN 37614 USA
[7] Univ Cincinnati, Coll Med, Cincinnati, OH 45221 USA
关键词
bone mineral density; dual-energy X-ray absorptiometry; DXA; precision error; T-score;
D O I
10.1016/j.jocd.2007.03.002
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
We quantified confidence intervals (CIs) for T-scores for the lumbar spine and hip and determined the practical effect (impact on diagnosis) of variability around the T-score cutpoint of -2.5. Using precision data from the literature for GE Lunar Prodigy dual-energy X-ray absorptiometry (DXA) systems, the 95% CI for the T-score was +/- 0.23 at the lumbar spine (L1-L4), +/- 0.20 at the total hip, and +/- 0.41 at the femoral neck. Thus, T-score variations of +/- 0.23 or less at the spine, +/- 0.20 at the total hip, and +/- 0.41 at the femoral neck are not statistically significant. When diagnosing osteoporosis, T-scores in the interval -2.3 to -2.7 for spine or total hip (after rounding to conform to guidelines from the International Society for Clinical Densitometry) and -2.1 to -2.9 for femoral neck are not statistically different from -2.5. Better precision values resulted in smaller 95% CIs. This concept was applied to actual clinical data using Hologic DXA systems. The study cohort comprised 2388 white women with either normal or osteopenic spines in whom the densitometric diagnosis of osteoporosis would be determined by hip T-scores. When evaluating actual patient T-scores in the range -2.5 +/- 95% CI, we found that the diagnosis was indeterminate in approximately 12% of women when T-scores for femoral neck were used and in 4% of women when T-scores for total hip were used, with uncertainty as to whether the classification was osteopenia or osteoporosis. We conclude that precision influences the variability around T-scores and that this variability affects the reliability of diagnostic classification.
引用
收藏
页码:239 / 243
页数:5
相关论文
共 15 条
[1]
Importance of precision in bone density measurements [J].
Bonnick, SL ;
Johnston, CC ;
Kleerekoper, M ;
Lindsay, R ;
Miller, P ;
Sherwood, L ;
Siris, E .
JOURNAL OF CLINICAL DENSITOMETRY, 2001, 4 (02) :105-110
[2]
The tale of the T-score: review and perspective [J].
Faulkner, KG .
OSTEOPOROSIS INTERNATIONAL, 2005, 16 (04) :347-352
[3]
Faulkner KG, 2006, J CLIN DENSITOM, V9, P237, DOI DOI 10.1016/j.jocd.2006.04.045
[4]
FAVUS MJ, PRIMER METABOLIC BON
[5]
ACCURATE ASSESSMENT OF PRECISION ERRORS - HOW TO MEASURE THE REPRODUCIBILITY OF BONE DENSITOMETRY TECHNIQUES [J].
GLUER, CC ;
BLAKE, G ;
LU, Y ;
BLUNT, BA ;
JERGAS, M ;
GENANT, HK .
OSTEOPOROSIS INTERNATIONAL, 1995, 5 (04) :262-270
[6]
The prevalence of significant left-right differences in hip bone mineral density [J].
Hamdy, R. ;
Kiebzak, G. M. ;
Seier, E. ;
Watts, N. B. .
OSTEOPOROSIS INTERNATIONAL, 2006, 17 (12) :1772-1780
[7]
Hamdy Ronald C, 2002, J Clin Densitom, V5 Suppl, pS11, DOI 10.1385/JCD:5:3S:S11
[8]
Skeletal sites for osteoporosis diagnosis: The 2005 ISCD official positions [J].
Hans, Didier ;
Downs, Robert W., Jr. ;
Duboeuf, Francois ;
Greenspan, Susan ;
Jankowski, Lawrence G. ;
Kiebzak, Gary M. ;
Petak, Steven M. .
JOURNAL OF CLINICAL DENSITOMETRY, 2006, 9 (01) :15-21
[9]
Relationship among MRTA, DXA, and QUS revisited [J].
Kiebzak, GM ;
Ambrose, CG .
JOURNAL OF CLINICAL DENSITOMETRY, 2005, 8 (04) :396-403
[10]
Leib E, 2004, J CLIN DENSITOM, V7, P45