摘要
Background: Bearing surfaces in total hip arthroplasty (THA) may affect implant longevity and hence patient outcomes. This randomized clinical trial determined how ceramic-on-ceramic (CERAMIC) bearing THA affected joint-specific pain, function, and stiffness and prosthesis fixation/longevity over 10 postoperative years compared with ceramic-on-highly-crosslinked-polyethylene (POLYETHYLENE) bearing THA. This is a follow-up to previously reported 5-year outcomes. Methods: Subjects aged <61 years were randomized to CERAMIC (n = 48) or POLYETHYLENE (n = 44) THA. Subjects were assessed using the Western Ontario McMaster Osteoarthritis Index and the RAND 12-Item Health Survey preoperatively, and at 1, 5, and 10 years postoperatively. Plain radiographs were evaluated at 10 years for fixation, and medical records were reviewed for revisions. Results: Of 92 subjects, 6 (7%) died within 10 years; 68 (79%) survivors provided radiographic and/or clinical follow-up at 10 years postoperatively. Improvements seen at 5 years in both the Western Ontario McMaster Osteoarthritis Index and RAND 12-Item Health Survey were retained at 10 years with no group differences (P > .48). There were no failures or loss of fixation related to bearing surfaces/wear in either group. Over 10 years, 3 subjects in the POLYETHYLENE group had revisions that were related to recurrent dislocation; 2 revisions were performed within 2 years of surgery, and one further subject underwent revision at 7 years postoperatively. Conclusions: This is one of the first randomized clinical trials to examine 10-year outcomes between CERAMIC and POLYETHYLENE bearing THAs. Both bearing surfaces performed well out to 10 years in subjects who were <61 years at time of surgery. (C) 2015 Elsevier Inc. All rights reserved.
引用
收藏
页码:1240 / 1245
页数:6
相关论文
共 16 条
[1]
Comparison of Surgical Outcomes and Implant Wear Between Ceramic-Ceramic and Ceramic-Polyethylene Articulations in Total Hip Arthroplasty [J].
Amanatullah, Derek F. ;
Landa, Joshua ;
Strauss, Eric J. ;
Garino, Jonathan P. ;
Kim, Sunny H. ;
Di Cesare, Paul E. .
JOURNAL OF ARTHROPLASTY, 2011, 26 (06) :72-77
[2]
Alumina-on-alumina ceramic versus metal-on-highly cross-linked polyethylene bearings in total hip arthroplasty: a comparative study [J].
Bascarevic, Zoran ;
Vukasinovic, Zoran ;
Slavkovic, Nemanja ;
Dulic, Borislav ;
Trajkovic, Goran ;
Bascarevic, Violeta ;
Timotijevic, Sladjan .
INTERNATIONAL ORTHOPAEDICS, 2010, 34 (08) :1129-1135
[3]
A Randomized Trial Of Ceramic-On-Ceramic Bearing Versus Ceramic-On-Crossfire-Polyethylene Bearing In Total Hip Arthroplasty: Five-Year Outcomes [J].
Beaupre, Lauren A. ;
Manolescu, Andrei ;
Johnston, D. W. C. .
JOURNAL OF ARTHROPLASTY, 2013, 28 (03) :485-489
[4]
Ceramic-on-ceramic bearings in young patients OUTCOMES AND ACTIVITY LEVELS AT MINIMUM TEN-YEAR FOLLOW-UP [J].
Chana, R. ;
Facek, M. ;
Tilley, S. ;
Walter, W. K. ;
Zicat, B. ;
Walter, W. L. .
BONE & JOINT JOURNAL, 2013, 95B (12) :1603-1609
[5]
Nine years follow-up of 100 ceramic-on-ceramic total hip arthroplasty [J].
Chevillotte, Christophe ;
Pibarot, Vincent ;
Carret, Jean-Paul ;
Bejui-Hugues, Jacques ;
Guyen, Olivier .
INTERNATIONAL ORTHOPAEDICS, 2011, 35 (11) :1599-1604
[6]
No Differences Found in Bearing Related Hip Survivorship at 10-12 Years Follow-Up Between Patients With Ceramic on Highly Cross-Linked Polyethylene Bearings Compared to Patients With Ceramic on Ceramic Bearings [J].
Epinette, Jean-Alain ;
Manley, Michael T. .
JOURNAL OF ARTHROPLASTY, 2014, 29 (07) :1369-1372
[7]
Alumina-on-alumina total hip arthroplasty - A minimum 18.5-year follow-up study [J].
Hamadouche, M ;
Boutin, P ;
Daussange, J ;
Bolander, ME ;
Sedel, L .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (01) :69-77
[8]
Ceramic-Ceramic Bearing Decreases Osteolysis: A 20-year Study versus Ceramic-Polyethylene on the Contralateral Hip [J].
Hernigou, Philippe ;
Zilber, Sebastien ;
Filippini, Paolo ;
Poignard, Alexandre .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (09) :2274-2280
[9]
Comparison of ceramic-on-ceramic to metal-on-polyethylene bearing surfaces in total hip arthroplasty: a meta-analysis of randomized controlled trials [J].
Hu, Dongcai ;
Tie, Kai ;
Yang, Xiao ;
Tan, Yang ;
Alaidaros, Mohammed ;
Chen, Liaobin .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2015, 10
[10]
Cementless total hip arthroplasty with ceramic-on-ceramic bearing in patients younger than 45 years with femoral-head osteonecrosis [J].
Kim, Young-Hoo ;
Choi, Yoowang ;
Kim, Jun-Shik .
INTERNATIONAL ORTHOPAEDICS, 2010, 34 (08) :1123-1127