Multifrequency ultrasound transducers for conformal interstitial thermal therapy

被引:47
作者
Chopra, R [1 ]
Luginbuhl, C
Foster, FS
Bronskill, MJ
机构
[1] Sunnybrook & Womens Coll, Hlth Sci Ctr, Toronto, ON M4N 3M5, Canada
[2] Univ Toronto, Dept Med Biophys, Toronto, ON M4N 3M5, Canada
关键词
D O I
10.1109/TUFFC.2003.1214507
中图分类号
O42 [声学];
学科分类号
070206 [声学]; 082403 [水声工程];
摘要
Control over the pattern of thermal damage generated by interstitial ultrasound heating applicators can be enhanced by changing the ultrasound frequency during heating. The ability to change transmission frequency from a single transducer through the use of high impedance front layers was investigated in this study. The transmission spectrum of multifrequency transducers was calculated using the KLM equivalent circuit model and verified with experimental measurements on prototype transducers. The addition of a quarter-wavelength thick PZT (unpoled) front layer enabled the transmission of ultrasound at two discrete frequencies, 4.7 and 9.7 MHz, from a transducer with an original resonant frequency of 8.4 MHz. Three frequency transmission at 3.3, 8.4, and 10.8 MHz was possible for a transducer with a half-wavelength thick front layer. Calculations of the predicted thermal lesion size at each transmission frequency indicated that the depth of thermal lesion could be varied by a factor of 1.6 for the quarter-wavelength front layer. Heating experiments performed in excised liver tissue with a dual-frequency applicator confirmed this ability to control the shape of thermal lesions during heating to generate a desired geometry. Practical interstitial designs that enable the generation of shaped thermal lesions are feasible.
引用
收藏
页码:881 / 889
页数:9
相关论文
共 25 条
[1]
Interstitial ultrasound heating applicator for MR-guided thermal therapy [J].
Chopra, R ;
Luginbuhl, C ;
Weymouth, AJ ;
Foster, FS ;
Bronskill, MJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (12) :3133-3145
[2]
Feasibility of linear arrays for interstitial ultrasound thermal therapy [J].
Chopra, R ;
Bronskill, MJ ;
Foster, FS .
MEDICAL PHYSICS, 2000, 27 (06) :1281-1286
[3]
Dependence of ultrasonic attenuation and absorption in dog soft tissues on temperature and thermal dose [J].
Damianou, CA ;
Sanghvi, NT ;
Fry, FJ ;
MaassMoreno, R .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1997, 102 (01) :628-634
[4]
Angular directivity of thermal coagulation using air-cooled direct-coupled interstitial ultrasound applicators [J].
Deardorff, DL ;
Diederich, CJ .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1999, 25 (04) :609-622
[5]
DESIGN OF EFFICIENT BROAD-BAND PIEZOELECTRIC TRANSDUCERS [J].
DESILETS, CS ;
FRASER, JD ;
KINO, GS .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1978, 25 (03) :115-125
[6]
Ultrasound applicators for interstitial thermal coagulation [J].
Diederich, CJ ;
Nau, WH ;
Stauffer, PR .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1999, 46 (05) :1218-1228
[7]
DESIGN OF BROAD-BAND FLUID-LOADED ULTRASONIC TRANSDUCERS [J].
GOLL, JH .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1979, 26 (06) :385-393
[8]
GUYTON A.C., 1981, TXB MED PHYSL
[9]
DESIGN OF ULTRASONIC TRANSDUCERS WITH MULTIPLE ACOUSTIC MATCHING LAYERS FOR MEDICAL APPLICATION [J].
INOUE, T ;
OHTA, M ;
TAKAHASHI, S .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1987, 34 (01) :8-16
[10]
EFFECTS OF BACKING AND MATCHING ON PERFORMANCE OF PIEZOELECTRIC CERAMIC TRANSDUCERS [J].
KOSSOFF, G .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1966, SU13 (01) :20-&