Solid-cantact electrochemical polyion sensors for monitoring peptidase activities

被引:25
作者
Fordyce, Katherine [1 ]
Shvarev, Alexey [1 ]
机构
[1] Oregon State Univ, Dept Chem, Corvallis, OR 97331 USA
关键词
D O I
10.1021/ac701775n
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We report here on improved solid-contact electrochemical polyion sensors for the detection of polyion protamine. The polymeric membrane sensors were fabricated with a conducting polymer as an ion-electron transduction layer. We observed. that decreasing the magnitude of the applied current pulse caused a significant improvement of the sensor sensitivity to low protamine levels. The protamine sensors exhibited a stable and reversible response to protamine concentrations ranging from 0.05 to 30 mg L-1. The sensors were used for monitoring peptidase activities utilizing galvanostatically controlled solid-contact membrane sensors. The polyion protamine was used as a substrate to detect the activity of the protease trypsin. The enzyme activity was continuously monitored by measuring the protamine concentration as it is cleaved by enzyme into smaller fragments to which the sensor is less sensitive. In the presence of a given level of protamine the initial rate of reaction can be linearly related to the trypsin activity within a 0-5 U mL(-1) concentration range. The interference with the enzymatic reaction product arginine was specifically examined.
引用
收藏
页码:827 / 833
页数:7
相关论文
共 23 条
[1]  
Berg J M., 2007, Biochemistry, V6
[2]  
Dai S, 2001, ELECTROANAL, V13, P276, DOI 10.1002/1521-4109(200103)13:4<276::AID-ELAN276>3.0.CO
[3]  
2-V
[4]   Influence of a water rinse on the structure and properties of poly(3,4-ethylene dioxythiophene):poly(styrene sulfonate) films [J].
DeLongchamp, DM ;
Vogt, BD ;
Brooks, CM ;
Kano, K ;
Obrzut, J ;
Richter, CA ;
Kirillov, OA ;
Lin, EK .
LANGMUIR, 2005, 21 (24) :11480-11483
[5]   RESPONSE MECHANISM OF POLYMER MEMBRANE-BASED POTENTIOMETRIC POLYION SENSORS [J].
FU, B ;
BAKKER, E ;
YUN, JH ;
YANG, VC ;
MEYERHOFF, ME .
ANALYTICAL CHEMISTRY, 1994, 66 (14) :2250-2259
[6]   Selective monitoring of peptidase activities with synthetic polypeptide substrates and polyion-sensitive membrane electrode detection [J].
Han, IS ;
Ramamurthy, N ;
Yun, JH ;
Schaller, U ;
Meyerhoff, ME ;
Yang, VC .
FASEB JOURNAL, 1996, 10 (14) :1621-1626
[7]  
HISHHAM SM, 2003, BIOSENS BIOELECTRON, V18, P229
[8]   STRUCTURAL BASIS OF ACTIVATION AND ACTION OF TRYPSIN [J].
HUBER, R ;
BODE, W .
ACCOUNTS OF CHEMICAL RESEARCH, 1978, 11 (03) :114-122
[9]  
Jadhav S, 2000, ELECTROANAL, V12, P1251, DOI 10.1002/1521-4109(200011)12:16<1251::AID-ELAN1251>3.0.CO
[10]  
2-P