Modes of wear after semicionstrained total elbow arthroplasty

被引:69
作者
Goldberg, Steven H. [1 ]
Urban, Robert M. [1 ]
Jacobs, Joshua J. [1 ]
King, Graham J. W. [1 ]
O'Driscoll, Shawn W. [1 ]
Cohen, Mark S. [1 ]
机构
[1] Rush Univ, Med Ctr, Dept Orthopaed Surg, Chicago, IL 60612 USA
关键词
D O I
10.2106/JBJS.F.01286
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Osteolysis and aseptic loosening are increasingly recognized complications of total elbow arthroplasty. However, unlike the literature on total hip and knee arthroplasty, studies describing the mechanisms of these processes after total elbow arthroplasty are sparse. Methods: Serniconstrained total elbow arthroplasty components were retrieved from sixteen elbows (fourteen patients) at either revision surgery (at a mean of five years after implantation) for mechanical failure (fifteen elbows) or postmortem examination (one elbow). In all cases, the retrieved implant was the primary implant. The patterns of damage on these components were investigated with stereomicroscopy in correlation with clinical findings, serial radiographs, and histopathological observations. Results: All of the retrieved devices exhibited multiple modes of wear. Damage to the humeral and ulnar polyethylene bushings was nearly universal; twenty-seven of twenty-eight humeral bushings demonstrated asymmetrical thinning, while fifteen of sixteen ulnar bushings demonstrated elliptical plastic deformation. In addition, unintended metal-on-metal wear between bearing and nonbearing surfaces or between two nonbearing surfaces was commonly observed, typically in association with wear and deformation of the polyethylene bushings. Wear between the stem and the cement mantle was observed in most of the ulnar components. The histopathology of the periprosthetic tissues was similar in character to that observed in association with osteolysis and loosening of total hip and knee replacements, while analysis of the particulate debris revealed a preponderance of titanium alloy and polyethylene debris. Barium sulfate particles were also observed to a lesser extent. Conclusions: Multimodal wear in total elbow replacements can lead to osteolysis, aseptic loosening, and prosthetic and periprosthetic fracture necessitating revision surgery. Polyethylene wear and damage, as well as unintended metal-on-metal wear, contribute to the periprosthetic particulate burden, which is likely pathogenic in these processes.
引用
收藏
页码:609 / 619
页数:11
相关论文
共 38 条
[1]
Aldridge JM, 2006, J BONE JOINT SURG BR, V88B, P509, DOI 10.1302/0301-620X.88B4
[2]
Revision total elbow arthroplasty for prosthetic fractures [J].
Athwal, George S. ;
Morrey, Bernard F. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (09) :2017-2026
[3]
Total elbow arthroplasty as primary treatment for distal humeral fractures in elderly patients [J].
Cobb, TK ;
Morrey, BF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (06) :826-832
[4]
Total elbow arthroplasty for treatment of elbow arthritis [J].
Ferlic, DC .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 1999, 8 (04) :367-378
[5]
Figgie H E 3rd, 1986, J Arthroplasty, V1, P169, DOI 10.1016/S0883-5403(86)80027-4
[6]
Locking mechanism failure in semiconstrained total elbow arthroplasty [J].
Figgie, MP ;
Su, EP ;
Kahn, B ;
Lipman, J .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2006, 15 (01) :88-93
[7]
TOTAL ELBOW ARTHROPLASTY FOR COMPLETE ANKYLOSIS OF THE ELBOW [J].
FIGGIE, MP ;
INGLIS, AE ;
MOW, CS ;
FIGGIE, HE .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1989, 71A (04) :513-520
[8]
FISHER J, 2006, JOINT REPLACEMENT BO, P145
[9]
The Coonrad-Morrey total elbow arthroplasty in patients who have rheumatoid arthritis - A ten to fifteen-year follow-up study [J].
Gill, DRJ ;
Morrey, BF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (09) :1327-1335
[10]
Hastings Hill 2nd, 2004, Tech Hand Up Extrem Surg, V8, P34, DOI 10.1097/00130911-200403000-00008