Low temperature thermometry in high magnetic fields.: VII.: Cernox™ sensors to 32 T

被引:105
作者
Brandt, BL [1 ]
Liu, DW
Rubin, LG
机构
[1] Natl High Magnet Field Lab, Tallahassee, FL 32310 USA
[2] MIT, Cambridge, MA 02139 USA
关键词
D O I
10.1063/1.1149549
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This article describes the behavior of Cernox(TM) zirconium oxynitride film temperature sensors from Lake Shore Cryotronics, Inc. in magnetic fields up to 32 T at temperatures between 2 and 286 K. Results from a number of sensors with different dimensionless temperature sensitivities and from different production batches are analyzed and compared with previous results on carbon-glass sensors that extend to 19 T. In that field range, the Cernox(TM) sensors appear to be a real alternative to carbon glass. Furthermore, the magnitude of their uncorrected error, Delta T/T, is smaller than other sensors at most temperatures in fields less than 20 T, and their temperature correctabilities appear to apply to off-the-shelf sensors with dimensionless temperature sensitivities in the range of -0.74 to -1.9. The sensors show a negligible (<0.05%) orientation dependence of their Delta R/R at 78 K; at 4.2 K, that dependence can be as high as similar to 0.7% at 20 T. (C) 1999 American Institute of Physics. [S0034-6748(99)04701-2].
引用
收藏
页码:104 / 110
页数:7
相关论文
共 12 条
[11]   NEW TYPE CRYOGENIC THERMOMETER USING SPUTTERED ZR-N FILMS [J].
YOTSUYA, T ;
YOSHITAKE, M ;
YAMAMOTO, J .
APPLIED PHYSICS LETTERS, 1987, 51 (04) :235-237
[12]   GERMANIUM RESISTANCE THERMOMETERS WITH LOW MAGNETORESISTANCE [J].
ZARUBIN, LI ;
NEMISH, IY ;
SZMYRKAGRZEBYK, A .
CRYOGENICS, 1990, 30 (06) :533-537