Quality of motion considerations in numerical analysis of motion restoring implants of the spine

被引:35
作者
Bowden, Anton E. [1 ]
Guerin, Heather L. [1 ]
Villarraga, Marta L. [1 ]
Patwardhan, Avinash G. [2 ]
Ochoa, Jorge A. [3 ]
机构
[1] Exponent Inc, Philadelphia, PA 19104 USA
[2] Loyola Univ, Med Ctr, Maywood, IL 60153 USA
[3] Archus Orthoped, Redmond, WA USA
关键词
quality of motion; spine; finite element analysis; kinematic signature;
D O I
10.1016/j.clinbiomech.2007.12.010
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Background. Motion restoring implants function in a dynamic environment that encompasses the full range of spinal kinematics. Accurate assessment of the in situ performance of these devices using numerical techniques requires model verification and validation against the well-established nonlinear quality of motion of the spine, as opposed to the previous norm of matching kinematic endpoint metrics such as range of motion and intervertebral disc pressure measurements at a single kinematic reference point. Methods. Experimental data was obtained during cadaveric testing of nine three-functional spinal unit (L3-S1 ) lumbar spine segments. Each specimen was tested from 8 Nm of applied flexion moment to 6 Nm of applied extension moment with an applied 400 N compressive follower preload. A nonlinear kinematic curve representing the spinal quality of motion (applied moment versus angular rotation) for the index finite element model was constructed and compared to the kinematic responses of the experimental specimens. The effect of spinal soft tissue structure mechanical behaviors on the fidelity of the model's quality of motion to experimental data was assessed by iteratively modifying the material representations of annulus fibrosus, nucleus pulposus, and ligaments. Findings. The present work demonstrated that for this model, the annulus fibrosus played a small role in the nonlinear quality of motion of the model, whereas changes in ligament representations had a large effect, as validated against the full kinematic range of motion. An anisotropic continuum representation of the annulus fibrosus was used, along with nonlinear fabric representations of the ligaments and a hyperelastic representation of the nucleus pulposus. Interpretation. Our results suggest that improvements in current methodologies broadly used in numerical simulations of the lumbar spine are needed to fully describe the highly nonlinear motion of the spine. (c) 2007 Elsevier Ltd. All rights reserved.
引用
收藏
页码:536 / 544
页数:9
相关论文
共 39 条
[11]
Eberlein R., 2001, COMPUT METHOD BIOMEC, V4, P209, DOI [10.1080/10255840108908005, DOI 10.1080/10255840108908005]
[12]
Anisotropic and inhomogeneous tensile behavior of the human anulus fibrosus: Experimental measurement and material model predictions [J].
Elliott, DM ;
Setton, LA .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2001, 123 (03) :256-263
[13]
The effect of nucleotomy on lumbar spine mechanics in compression and shear loading [J].
Frei, H ;
Oxland, TR ;
Rathonyi, GC ;
Nolte, LP .
SPINE, 2001, 26 (19) :2080-2089
[14]
GARDNERMORSE MG, 2003, 49 ANN M ORTH RES SO
[15]
INTERLAMINAR SHEAR STRESSES AND LAMINAE SEPARATION IN A DISC - FINITE-ELEMENT ANALYSIS OF THE L3-L4 MOTION SEGMENT SUBJECTED TO AXIAL COMPRESSIVE LOADS [J].
GOEL, VK ;
MONROE, BT ;
GILBERTSON, LG ;
BRINCKMANN, P ;
NAT, R .
SPINE, 1995, 20 (06) :689-698
[16]
Validation of a clinical finite element model of the human lumbosacral spine [J].
Guan, Yabo ;
Yoganandan, Narayan ;
Zhang, Jiangyue ;
Pintar, Frank A. ;
Cusick, Joesph F. ;
Wolfla, Christopher E. ;
Maiman, Dennis J. .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2006, 44 (08) :633-641
[17]
Degeneration affects the fiber reorientation of human annulus fibrosus under tensile load [J].
Guerin, Heather Anne L. ;
Elliott, Dawn M. .
JOURNAL OF BIOMECHANICS, 2006, 39 (08) :1410-1418
[18]
Ouantifying the contributions of structure to annulus fibrosus mechanical function using a nonlinear, anisotropic, hyperelastic model [J].
Guerin, Heather Lynch ;
Elliott, Dawn M. .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2007, 25 (04) :508-516
[19]
Hallquist J., 2006, LS DYNA THEORY MANUA
[20]
Role of facet curvature for accurate vertebral facet load analysis [J].
Holzapfel, Gerhard A. ;
Stadler, Michael .
EUROPEAN SPINE JOURNAL, 2006, 15 (06) :849-856