Polyurethane foam as a sorbent for continuous flow analysis: Preconcentration and spectrophotometric determination of zinc in biological materials

被引:65
作者
de Jesus, DS
Casella, RJ
Ferreira, SLC
Costa, ACS
de Carvalho, MS
Santelli, RE [1 ]
机构
[1] Univ Fed Fluminense, Dept Geochem, BR-24020007 Niteroi, RJ, Brazil
[2] Fed Bahia Ctr Technol Educ, Salvador, BA, Brazil
[3] Fed Sch Tech Chem, Rio De Janeiro, Brazil
[4] Univ Fed Bahia, Inst Chem, Dept Analyt Chem, BR-40170290 Salvador, BA, Brazil
[5] CNEN, IEN, Nucl Engineer Inst, Rio De Janeiro, Brazil
[6] Univ Fed Fluminense, Dept Geochem, BR-24020007 Niteroi, RJ, Brazil
关键词
polyurethane foam; zinc preconcentration; continuous flow analysis; biological samples;
D O I
10.1016/S0003-2670(98)00111-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple and accurate procedure was developed to preconcentrate and determine zinc in biological matrices. A polyurethane foam (PUF) minicolumn was used for the first time for on-line preconcentration. This system was applied to preconcentrate zinc from aqueous solutions which is sorbed as zinc-thiocyanate complex. Then, the complex was eluted by water and zinc was determined by spectrophotometry using 4-(2-pyridylazo)-resorcinol. The method was validated by the analysis of several biological certified reference materials. The detection limit was 0.9 ng ml(-1) and the RSD was 1.2% for 1 min preconcentration time. Generally, concentration of Fe(III), Al(III), Mo(VI), Ni(II), Mn(II), Cu(II), and Co(II) found in biological matrices did not interfere. Dynamic ranges of concentration from 20 to 100 ng ml(-1) (1 min preconcentration time) and from 5 to 20 ng ml(-1) (5 min preconcentration time) achieve a sample throughput of 20 and 8 samples per hour and a preconcentration factor of 8 and 32, respectively. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:263 / 269
页数:7
相关论文
共 35 条
[1]   COLUMN SOLID-PHASE EXTRACTION AS PRECONCENTRATION METHOD FOR TRACE-ELEMENT DETERMINATION IN OXALIC-ACID BY ATOMIC-ABSORPTION SPECTROMETRY AND INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
ALEXANDROVA, A ;
ARPADJAN, S .
ANALYTICA CHIMICA ACTA, 1995, 307 (01) :71-77
[2]   Luminescence determination of europium microquantities after its preconcentration on polyurethane foam [J].
Beltyukova, S ;
Balamtsarashvili, C .
TALANTA, 1995, 42 (12) :1833-1838
[3]   SORPTION OF YTTRIUM HYDROXYQUINOLINATES BY POLYURETHANE FOAM AND ITS USE IN ROCK ANALYSIS [J].
BELTYUKOVA, SV ;
NAZARENKO, NA ;
TSYGANKOVA, S .
ANALYST, 1995, 120 (06) :1693-1698
[4]   ABSORPTION BY POLYURETHANE FOAMS - NEW METHOD OF SEPARATION [J].
BOWEN, HJM .
JOURNAL OF THE CHEMICAL SOCIETY A -INORGANIC PHYSICAL THEORETICAL, 1970, (07) :1082-&
[5]   CHROMOFOAMS QUALITATIVE AND SEMI-QUANTITATIVE TESTS WITH CHROMOGENIC ORGANIC REAGENTS IMMOBILIZED IN PLASTICIZED OPEN-CELL POLYURETHANE FOAMS [J].
BRAUN, T ;
FARAG, AB .
ANALYTICA CHIMICA ACTA, 1974, 73 (02) :301-309
[6]   POLYURETHANE FOAMS AND MICROSPHERES IN ANALYTICAL-CHEMISTRY - IMPROVED LIQUID-SOLID, GAS-SOLID AND LIQUID-LIQUID CONTACT VIA A NEW GEOMETRY OF SOLID-PHASE [J].
BRAUN, T ;
FARAG, AB .
ANALYTICA CHIMICA ACTA, 1978, 99 (01) :1-36
[7]   CELLULAR AND FOAMED PLASTICS AS SEPARATION MEDIA - NEW GEOMETRICAL FORM OF SOLID-PHASE IN ANALYTICAL LIQUID-SOLID CONTACT [J].
BRAUN, T ;
FARAG, AB .
TALANTA, 1975, 22 (09) :699-705
[8]  
Braun T., 1985, POLYURETHANE FOAM SO
[9]   Selective determination of cobalt using polyurethane foam and 2-(2-benzothiazolylazo)-2-p-cresol as a spectrophotometric reagent [J].
Carvalho, MS ;
Fraga, ICS ;
Neto, KCM ;
Silva, EQ .
TALANTA, 1996, 43 (10) :1675-1680
[10]  
DEJESUS DS, UNPUB TALANTA