Uterine arterial embolization: Factors influencing patient radiation exposure

被引:59
作者
Andrews, RT [1 ]
Brown, PH [1 ]
机构
[1] Oregon Hlth Sci Univ, Dotter Intervent Inst, Portland, OR 97201 USA
关键词
arteries; therapeutic embolization; dosimetry; leiomyoma; radiations; exposure to patients and personnel; uterine neoplasms; therapy;
D O I
10.1148/radiology.217.3.r00dc30713
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To investigate patient radiation exposures during uterine arterial embolization and the factors responsible for those exposures. MATERIALS AND METHODS: Clinical and procedural factors were evaluated for 42 consecutive procedures performed in 39 patients by one operator. Seven patients were excluded because of early termination (n = 1) or unusual conditions that necessitated extended procedures (n = 6). Fluoroscopic time, number of images acquired, height, and weight were available in the 35 remaining patients, and dose-area product (DAP) was available in 20. Equipment factors were evaluated by using a Lucite phantom in four angiography units from three manufacturers. RESULTS: The mean fluoroscopic time per case decreased from 30.6 to 14.2 minutes between the 1st and 5th quintiles. Mean DAP decreased from 211.4 to 30.6 Gy(.)cm(2) with dose reduction techniques; this primarily reflected a decreased number of acquired images. Phantom studies demonstrated many significant dose variations with magnification and equipment position. Low-dose and pulsed fluoroscopic modes reduced exposure rates in units so equipped, but roadmapping caused a silent switch to continuous fluoroscopy in two such units, which doubled the exposure rate. CONCLUSION: With operator experience and careful technique, uterine arterial embolization can be performed at radiation exposures comparable to those used in routine diagnostic studies. However, operators must be familiar with the technical parameters of their angiographic equipment.
引用
收藏
页码:713 / 722
页数:10
相关论文
共 22 条
[11]   Selective uterine artery embolization as primary treatment for symptomatic leiomyomata uteri [J].
Hutchins, FL ;
Worthington-Kirsch, R ;
Berkowitz, RP .
JOURNAL OF THE AMERICAN ASSOCIATION OF GYNECOLOGIC LAPAROSCOPISTS, 1999, 6 (03) :279-284
[12]  
LEHERON JC, 1992, PHYS MED BIOL, V37, P2117, DOI 10.1088/0031-9155/37/11/008
[13]   Patient radiation dose associated with uterine artery embolization [J].
Nikolic, B ;
Spies, JB ;
Lundsten, MJ ;
Abbara, S .
RADIOLOGY, 2000, 214 (01) :121-125
[14]  
NIKOLIC B, 1999, 24 ANN M SOC CARD IN, P183
[15]  
Pitman A G, 1997, Australas Radiol, V41, P329, DOI 10.1111/j.1440-1673.1997.tb00642.x
[16]   ARTERIAL EMBOLIZATION TO TREAT UTERINE MYOMATA [J].
RAVINA, JH ;
HERBRETEAU, D ;
CIRARUVIGNERON, N ;
BOURET, JM ;
HOUDART, E ;
AYMARD, A ;
MERLAND, JJ .
LANCET, 1995, 346 (8976) :671-672
[17]  
*SOC CARD INT RAD, 1999, UAE SURV RES 4000 PR
[18]   Initial results from uterine fibroid embolization for symptomatic leiomyomata [J].
Spies, JB ;
Scialli, AR ;
Jha, RC ;
Imaoka, I ;
Ascher, SM ;
Fraga, VM ;
Barth, KH .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1999, 10 (09) :1149-1157
[19]   Management of patient skin dose in fluoroscopically guided interventional procedures [J].
Wagner, LK ;
Archer, BR ;
Cohen, AM .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (01) :25-33
[20]   Radiation effective doses to patients undergoing abdominal CT examinations [J].
Ware, DE ;
Huda, W ;
Mergo, PJ ;
Litwiller, AL .
RADIOLOGY, 1999, 210 (03) :645-650