A model for the relative humidity effect on the readings of the PM10 beta-gauge monitor

被引:39
作者
Chang, CT [1 ]
Tsai, CJ [1 ]
机构
[1] Natl Chiao Tung Univ, Inst Environm Engn, Hsinchu 300, Taiwan
关键词
beta-gauge; PM10; aerosol mass monitor; humidity effect;
D O I
10.1016/S0021-8502(03)00356-2
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
In the previous paper (Atmos. Environ. 15 (1981) 1087), we found that the PM10 concentrations detected by the Wedding beta-gauge PM10 monitor and those measured by the manual hi-vol PM10 sampler were quite close when the ambient relative humidity (RH) was lower than the deliquescence RH (DRH) of aerosols. However, when the deliquescent point was exceeded, PM10 concentrations of the beta-gauge were found to be higher and differences increased with an increasing ambient RH. In addition, theoretical water mass calculated by a thermodynamic model (ISORROPIA model, (Aquat. Geochem. 4 (1998) 123)) was found to be much higher than the actual values. In this study, models were developed to determine water evaporation loss from collected particles on the filter tape of the beta-gauge during sampling and in the monitoring room. Simulated results show that all absorbed water will evaporate completely at RH lower than about 85%. However, absorbed water does not evaporate completely at RE higher than about 85%, and remaining water in particles accounts for higher beta-gauge readings than the hi-vol concentrations. The simulated daily beta-gauge PM10 concentrations are close to the actual beta-gauge readings obtained previously. (C) 2003 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1685 / 1697
页数:13
相关论文
共 11 条
[1]  
[Anonymous], 1987, FED REGISTER, V52, P24634
[2]   ATMOSPHERIC PARTICULATE NITRATE SAMPLING ERRORS DUE TO REACTIONS WITH PARTICULATE AND GASEOUS STRONG ACIDS [J].
APPEL, BR ;
TOKIWA, Y .
ATMOSPHERIC ENVIRONMENT, 1981, 15 (06) :1087-1089
[3]   Differences in PM10 concentrations measured by β-gauge monitor and hi-vol sampler [J].
Chang, CT ;
Tsai, CJ ;
Lee, CT ;
Chang, SY ;
Cheng, MT ;
Chein, HM .
ATMOSPHERIC ENVIRONMENT, 2001, 35 (33) :5741-5748
[4]   Evaporation loss of ammonium nitrate particles during filter sampling [J].
Cheng, YH ;
Tsai, CJ .
JOURNAL OF AEROSOL SCIENCE, 1997, 28 (08) :1553-1567
[5]   EFFECT OF HUMIDITY AND PARTICLE HYGROSCOPICITY ON THE MASS LOADING CAPACITY OF HIGH-EFFICIENCY PARTICULATE AIR (HEPA) FILTERS [J].
GUPTA, A ;
NOVICK, VJ ;
BISWAS, P ;
MONSON, PR .
AEROSOL SCIENCE AND TECHNOLOGY, 1993, 19 (01) :94-107
[6]  
MCFARLAND AR, 1987, 4716010881ARM TEX A, P15
[7]   ISORROPIA: A new thermodynamic equilibrium model for multiphase multicomponent inorganic aerosols [J].
Nenes, A ;
Pandis, SN ;
Pilinis, C .
AQUATIC GEOCHEMISTRY, 1998, 4 (01) :123-152
[8]   WIND-TUNNEL EVALUATION OF PM10 SAMPLERS [J].
RANADE, MB ;
WOODS, MC ;
CHEN, FL ;
PURDUE, LJ ;
REHME, KA .
AEROSOL SCIENCE AND TECHNOLOGY, 1990, 13 (01) :54-71
[9]  
WEDDING JB, 1993, J AIR WASTE MANAGE, V43, P475
[10]   THEORETICAL-ANALYSIS OF EVAPORATIVE LOSSES FROM IMPACTOR AND FILTER DEPOSITS [J].
ZHANG, XQ ;
MCMURRY, PH .
ATMOSPHERIC ENVIRONMENT, 1987, 21 (08) :1779-1789