Hydrodynamics-induced variability in the USP apparatus II dissolution test

被引:99
作者
Baxter, JL [1 ]
Kukura, J [1 ]
Muzzio, FJ [1 ]
机构
[1] Rutgers State Univ, Dept Chem & Biochem Engn, Piscataway, NJ 08854 USA
关键词
dissolution testing; hydrodynamics; USP apparatus II; shear; computational fluid dynamics;
D O I
10.1016/j.ijpharm.2004.08.003
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
The USP tablet dissolution test is an analytical tool used for the verification of drug release processes and formulation selection within the pharmaceutical industry. Given the strong impact of this test, it is surprising that operating conditions and testing devices have been selected empirically. In fact, the flow phenomena in the USP test have received little attention in the past. An examination of the hydrodynamics in the USP apparatus II shows that the device is highly vulnerable to mixing problems that can affect testing performance and consistency. Experimental and computational techniques reveal that the flow field within the device is not uniform, and dissolution results can vary dramatically with the position of the tablet within the vessel. Specifically, computations predict sharp variations in the shear along the bottom of the vessel where the tablet is most likely to settle. Experiments in which the tablet location was carefully controlled reveal that the variation of shear within the testing device can affect the measured dissolution rate. (c) 2004 Published by Elsevier B.V.
引用
收藏
页码:17 / 28
页数:12
相关论文
共 25 条
[1]  
Ballard J., 2003, Dissolution Technology, V10, P6, DOI DOI 10.14227/DTL00303P6
[2]  
BARDINA JE, 1997, TUBULENCE MODELING V
[3]   FLUID AND PARTICLE LASER DOPPLER VELOCITY-MEASUREMENTS AND MASS-TRANSFER PREDICTIONS FOR THE USP PADDLE METHOD DISSOLUTION APPARATUS [J].
BOCANEGRA, LM ;
MORRIS, GJ ;
JUREWICZ, JT ;
MAUGER, JW .
DRUG DEVELOPMENT AND INDUSTRIAL PHARMACY, 1990, 16 (09) :1441-1464
[4]  
Cox James C, 1982, RES EXPT EC, V2, P1
[5]  
*FDC, 2001, REC PROMPT FDA INCR, P1
[6]  
*FDC, 2003, SPIK POT REL PROBL C, P1
[7]  
*FDC, 2002, COUNT POS SPEC REC C, P1
[8]   LOS OF SENSITIVITY IN DISTINGUISHING REAL DIFFERENCES IN DISSOLUTION RATES DUE TO INCREASING INTENSITY OF AGITATION [J].
HAMLIN, WE ;
BALLARD, BE ;
WAGNER, JG ;
NELSON, E .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1962, 51 (05) :432-&
[9]   Sensitivity of dissolution rate to location in the paddle dissolution apparatus [J].
Healy, AM ;
McCarthy, LG ;
Gallagher, KM ;
Corrigan, OI .
JOURNAL OF PHARMACY AND PHARMACOLOGY, 2002, 54 (03) :441-444
[10]  
JOHNSON K, 2002, P 10 ANN C CFD SOC C, P490