Objectives: Breaking the skin when applying scalp electroencephalographic (EEG) electrodes creates the risk of infection from blood-born pathogens such as HIV, Hepatitis-C, and Creutzfeldt-Jacob Disease. Modern engineering principles suggest that excellent EEG signals can be collected with high scalp impedance (approximate to 40 k Ohm) without scalp abrasion. The present study was designed to evaluate the effect of electrode-scalp impedance on EEG data quality. Methods: The first section of the paper reviews electrophysiological recording with modern high input-impedance differential amplifiers and subject isolation, and explains how scalp-electrode impedance influences EEG signal amplitude and power line noise. The second section of the paper presents an experimental study of EEG data quality as a function of scalp-electrode impedance for the standard frequency bands in EEG and event-related potential (ERP) recordings and for 60 Hz noise. Results: There was no significant amplitude change in any EEG frequency bands as scalp-electrode impedance increased from less than 10 k Ohm (abraded skin) to 40 k Ohm (intact skin). 60 Hz was nearly independent of impedance mismatch, suggesting that capacitively coupled noise appearing differentially across mismatched electrode impedances did not contribute substantially to the observed 60 Hz noise levels. Conclusions: With modern high input-impedance amplifiers and accurate digital filters for power line noise, high-quality EEG can be recorded without skin abrasion. (C) 2001 Elsevier Science Ireland Ltd, All rights reserved.
引用
收藏
页码:536 / 544
页数:9
相关论文
共 13 条
[1]
[Anonymous], 1994, J Clin Neurophysiol, V11, P128
[2]
*CDC, 2000, INF CONTR DENT
[3]
*CDC, 1991, REC PREV TRANSM HUM
[4]
Ferree TC, 2000, IEEE T BIO-MED ENG, V47, P1584, DOI 10.1109/10.887939
[5]
TRANSMISSION OF CREUTZFELDT-JAKOB-DISEASE TO A CHIMPANZEE BY ELECTRODES CONTAMINATED DURING NEUROSURGERY [J].
GIBBS, CJ ;
ASHER, DM ;
KOBRINE, A ;
AMYX, HL ;
SULIMA, MP ;
GAJDUSEK, DC .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1994, 57 (06) :757-758
[6]
60-HZ INTERFERENCE IN ELECTROCARDIOGRAPHY [J].
HUHTA, JC ;
WEBSTER, JG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1973, BM20 (02) :91-101
[7]
Nunez PL, 1981, ELECT FIELDS BRAIN N
[8]
Guidelines for using human event-related potentials to study cognition: Recording standards and publication criteria [J].
Picton, TW ;
Bentin, S ;
Berg, P ;
Donchin, E ;
Hillyard, SA ;
Johnson, R ;
Miller, GA ;
Ritter, W ;
Ruchkin, DS ;
Rugg, MD ;
Taylor, MJ .
PSYCHOPHYSIOLOGY, 2000, 37 (02) :127-152
[9]
PUTNAM LE, 1992, PSYCHOPHYSIOLOGY, V29, P127
[10]
EEG ELECTRODE SENSITIVITY - AN APPLICATION OF RECIPROCITY [J].
RUSH, S ;
DRISCOLL, DA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1969, BM16 (01) :15-&