Cutting artefacts and cutting process in vitreous sections for cryo-electron microscopy

被引:191
作者
Al-Amoudi, A
Studer, D
Dubochet, J
机构
[1] Univ Lausanne, Lab Analyse Ultrastruct, CH-1015 Lausanne, Switzerland
[2] Univ Bern, Dept Anat, Inst Anat, CH-3000 Bern, Switzerland
关键词
CEMOVIS; knife marks; compression; crevasses; chatter; cutting artefacts;
D O I
10.1016/j.jsb.2005.01.003
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Cryo-electron microscopy of vitreous sections (CEMOVIS) has recently been shown to provide images of biological specimens with unprecedented quality and resolution. Cutting the sections remains however the major difficulty. Here, we examine the parameters influencing the quality of the sections and analyse the resulting artefacts. They are in particular: knife marks, compression, crevasses, and chatter. We propose a model taking into account the interplay between viscous flow and fracture. We confirm that crevasses are formed on only one side of the section, and define conditions by which they can be avoided. Chatter is an effect of irregular compression due to friction of the section of the knife edge and conditions to prevent this are also explored. In absence of crevasses and chatter, the bulk of the section is compressed approximately homogeneously. Within this approximation, it is possible to correct for compression by a simple linear transformation for the bulk of the section. A research program is proposed to test and refine our understanding of the sectioning process. (c) 2005 Elsevier Inc. All rights reserved.
引用
收藏
页码:109 / 121
页数:13
相关论文
共 35 条
[1]   An oscillating cryo-knife reduces cutting-induced deformation of vitreous ultrathin sections [J].
Al-Amoudi, A ;
Dubochet, J ;
Gnaegi, H ;
Lüthi, W ;
Studer, D .
JOURNAL OF MICROSCOPY, 2003, 212 :26-33
[2]   Amorphous solid water produced by cryosectioning of crystalline ice at 113 K [J].
Al-Amoudi, A ;
Dubochet, J ;
Studer, D .
JOURNAL OF MICROSCOPY, 2002, 207 (02) :146-153
[3]   Cryo-electron microscopy of vitreous sections of native biological cells and tissues [J].
Al-Amoudi, A ;
Norlen, LPO ;
Dubochet, J .
JOURNAL OF STRUCTURAL BIOLOGY, 2004, 148 (01) :131-135
[4]   Cryo-electron microscopy of vitreous sections [J].
Al-Amoudi, A ;
Chang, JJ ;
Leforestier, A ;
McDowall, A ;
Salamin, LM ;
Norlén, LPO ;
Richter, K ;
Blanc, NS ;
Studer, D ;
Dubochet, J .
EMBO JOURNAL, 2004, 23 (18) :3583-3588
[5]  
ALAMOUDI A, 2005, IN PRESS J INVEST DE
[6]   Amorphous water [J].
Angell, CA .
ANNUAL REVIEW OF PHYSICAL CHEMISTRY, 2004, 55 :559-583
[7]   Electron beam-induced changes in vitreous sections of biological samples [J].
Blanc, NS ;
Studer, D ;
Ruhl, K ;
Dubochet, J .
JOURNAL OF MICROSCOPY, 1998, 192 :194-201
[8]   DNA in human and stallion spermatozoa forms local hexagonal packing with twist and many defects [J].
Blanc, NS ;
Senn, A ;
Leforestier, A ;
Livolant, F ;
Dubochet, J .
JOURNAL OF STRUCTURAL BIOLOGY, 2001, 134 (01) :76-81
[9]   FREEZING, SECTIONING AND OBSERVATION ARTIFACTS OF FROZEN HYDRATED SECTIONS FOR ELECTRON-MICROSCOPY [J].
CHANG, JJ ;
MCDOWALL, AW ;
LEPAULT, J ;
FREEMAN, R ;
WALTER, CA ;
DUBOCHET, J .
JOURNAL OF MICROSCOPY, 1983, 132 (OCT) :109-123
[10]  
Christensen R., 1982, THEORY VISCOELASTICI