Capillary preconditioning for analysis of anions using indirect UV detection in capillary zone electrophoresis -: Systematic investigation of alkaline and acid prerinsing techniques by designed experiments

被引:16
作者
Ehmann, T
Bächmann, K
Fabry, L
Rüfer, H
Serwe, M
Ross, G
Pahlke, S
Kotz, L
机构
[1] Wacker Siltron AG, Cent Res & Dev, Cent Analyt Labs, D-84489 Burghausen, Germany
[2] Tech Univ Darmstadt, Dept Inorgan & Analyt Chem, D-64287 Darmstadt, Germany
[3] Wacker Siltron AG, Qual Syst Dept, D-84489 Burghausen, Germany
[4] Hewlett Packard GmbH, Chem Anal Grp Europe, D-76337 Waldbronn, Germany
关键词
chemometrics; column preconditioning; capillary columns; inorganic anions; organic acids;
D O I
10.1016/S0021-9673(98)00438-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
It is widely accepted in capillary zone electrophoresis, that the use of alkaline prerinse procedures can improve the reproducibility of migration times and of corrected peak areas. In this study we present a systematic investigation of alkaline and acid preconditioning procedures for anion analysis using indirect UV detection by designed experiments according to the methodology of Taguchi. Four frequently used electroosmotic flow modifiers (diethylenetriamine, hexamethonium hydroxide, tetradecyltrimethylammonium hydroxide, and hexadimethrine hydroxide) were examined. The optimized procedures were evaluated with regard to the necessary preconditioning time compared to the analysis time. Furthermore, it was demonstrated that the optimized preconditioning technique could be applied to other indirect detection based CE systems. For all examined electrolyte systems, relative standard deviations below 0.5% for migration times and below 5% for corrected peak areas (n=20) were achieved using automated peak integration without further manual reprocessing. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:261 / 275
页数:15
相关论文
共 43 条