Analyses and applications of single scan dose profiles in computed tomography

被引:24
作者
Tsai, HY
Tung, CJ [1 ]
Huang, MH
Wan, YL
机构
[1] Natl Tsing Hua Univ, Dept Nucl Sci, Hsinchu, Taiwan
[2] Lin Kuo Chang Gung Mem Hosp, Div Diagnost Radiol 1, Kuei Shan Hsiang, Taiwan
关键词
computed tomography; CT dose; CTDI; MSAD;
D O I
10.1118/1.1590315
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Comprehensive analyses and measurements of computed tomography (CT) single-scan dose profiles were performed for several scanners and operating conditions. Measurements were made using two types of thermoluminescent dosimeters, LiF:Mg,Cu,P and CaSO4:Dy, and two CT dosimetry phantoms, head and body. Analyses of CT single-scan dose profiles were made in terms of a Gaussian function for primary radiation and a Lorentzian function for scattered radiation. This function was used to investigate several common descriptions of the CT dose, including the computed tomography dose index (CTDI) and the multiple scan average dose. The relative percentage of scatter versus primary radiation to the contribution of CTDI at the central and peripheral locations was determined and analyzed. The correlation between CTDI of thermoluminescent dosimeter measurements and pencil-shaped ionization chamber measurements was determined. A method for estimating organ dose from CT was developed and compared to organ-dose estimates from Monte Carlo simulations. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:1982 / 1989
页数:8
相关论文
共 22 条
[1]  
American Association of Physicists in Medicine, 1977, 1 AAPM
[2]  
[Anonymous], 1984, ABSORPTION IONIZING
[3]  
[Anonymous], 1996, IAEA SAF SER
[4]   CALCULATED ENERGY-DEPENDENCE OF SOME RTL AND RPL DETECTORS [J].
BASSI, P ;
BUSUOLI, G ;
RIMONDI, O .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1976, 27 (5-6) :291-305
[5]   Study on examinee's dose delivered in computed tomography [J].
Cheung, T ;
Cheng, QJ ;
Feng, DH ;
Stokes, MJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (03) :813-820
[6]   AVERAGE RADIATION-DOSE IN STANDARD CT EXAMINATIONS OF THE HEAD - RESULTS OF THE 1990 NEXT SURVEY [J].
CONWAY, BJ ;
MCCROHAN, JL ;
ANTONSEN, RG ;
RUETER, FG ;
SLAYTON, RJ ;
SULEIMAN, OH .
RADIOLOGY, 1992, 184 (01) :135-140
[7]  
CRISTY M, 1987, ORNLTM8381V17
[8]  
FOX E, 1995, TRANSFORMS NONLINEAR
[9]  
*INT EL COMM, 1994, 6122326 INT EL COM 2
[10]  
*INT EL COMM, 2001, 60601244 INT EL CO 2