Analyses and applications of single scan dose profiles in computed tomography

被引:24
作者
Tsai, HY
Tung, CJ [1 ]
Huang, MH
Wan, YL
机构
[1] Natl Tsing Hua Univ, Dept Nucl Sci, Hsinchu, Taiwan
[2] Lin Kuo Chang Gung Mem Hosp, Div Diagnost Radiol 1, Kuei Shan Hsiang, Taiwan
关键词
computed tomography; CT dose; CTDI; MSAD;
D O I
10.1118/1.1590315
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Comprehensive analyses and measurements of computed tomography (CT) single-scan dose profiles were performed for several scanners and operating conditions. Measurements were made using two types of thermoluminescent dosimeters, LiF:Mg,Cu,P and CaSO4:Dy, and two CT dosimetry phantoms, head and body. Analyses of CT single-scan dose profiles were made in terms of a Gaussian function for primary radiation and a Lorentzian function for scattered radiation. This function was used to investigate several common descriptions of the CT dose, including the computed tomography dose index (CTDI) and the multiple scan average dose. The relative percentage of scatter versus primary radiation to the contribution of CTDI at the central and peripheral locations was determined and analyzed. The correlation between CTDI of thermoluminescent dosimeter measurements and pencil-shaped ionization chamber measurements was determined. A method for estimating organ dose from CT was developed and compared to organ-dose estimates from Monte Carlo simulations. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:1982 / 1989
页数:8
相关论文
共 22 条
[11]   Alternative methods of obtaining the computed tomography dose index [J].
Knox, HH ;
Gagne, RM .
HEALTH PHYSICS, 1996, 71 (02) :219-224
[12]  
Mettler Jr FA, 2000, J RADIOL PROT, V204, P353, DOI DOI 10.1088/0952-4746/20/4/301
[13]  
*NAT RAD PROT BROA, 1991, R249 NRPB
[14]  
*NAT RAD PROT BROA, 1991, R250 NRPB
[15]  
*OFF OFF PUBL EUR, 1999, 16262 OFF OFF PUBL E
[16]   A METHOD FOR DESCRIBING THE DOSES DELIVERED BY TRANSMISSION-X-RAY COMPUTED-TOMOGRAPHY [J].
SHOPE, TB ;
GAGNE, RM ;
JOHNSON, GC .
MEDICAL PHYSICS, 1981, 8 (04) :488-495
[17]   RADIATION-DOSIMETRY SURVEY OF COMPUTED-TOMOGRAPHY SYSTEMS FROM 10 MANUFACTURERS [J].
SHOPE, TB ;
MORGAN, TJ ;
SHOWALTER, CK ;
PENTLOW, KS ;
ROTHENBERG, LN ;
WHITE, DR ;
SPELLER, RD .
BRITISH JOURNAL OF RADIOLOGY, 1982, 55 (649) :60-69
[18]   CT scanner dosimetry [J].
Shrimpton, PC ;
Edyvean, S .
BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (841) :1-3
[19]  
Vergara JCS, 1999, RADIAT PROT DOSIM, V85, P429, DOI 10.1093/oxfordjournals.rpd.a032889
[20]   The use of small-angle x-ray diffraction studies for the analysis of structural features in archaeological samples [J].
Wess, TJ ;
Drakopoulos, M ;
Snigirev, A ;
Wouters, J ;
Paris, O ;
Fratzl, P ;
Collins, M ;
Hiller, J ;
Nielsen, K .
ARCHAEOMETRY, 2001, 43 :117-129