Influence of the packing heterogeneity on the performance of liquid chromatography supports

被引:29
作者
Billen, J [1 ]
Gzil, P [1 ]
Vervoort, N [1 ]
Baron, GV [1 ]
Desmet, G [1 ]
机构
[1] Vrije Univ Brussels, Dept Chem Engn, B-1050 Brussels, Belgium
关键词
packing heterogeneity; band broadening; flow resistance; computational fluid dynamics;
D O I
10.1016/j.chroma.2004.10.042
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We report on a series of plate height and flow resistance data obtained via computational fluid dynamics simulations in a simplified two-dimensional (2D) mimic of real packed bed and monolithic columns. By varying the external porosity (0.4 < ε < 0.8) and the degree of packing randomness, a good qualitative insight in the relationship between the packing porosity and heterogeneity and the general chromatographic performance parameters is obtained, unbiased by any differences in phase retention factor k', mobile phase diffusivity or viscosity or intra-skeleton porosity. The results provide a quantitative support for the use of domain size reduced plate heights as a means to compare the performance of chromatographic beds with a different porosity, as it was found that packings with a similar degree of packing heterogeneity yield very similar domain size reduced h(min)-values, nearly completely independent of the porosity. The study also clearly shows that the presence of preferential flow paths (inevitably accompanied by the presence of more clustered regions) leads to a decrease of the flow resistance, but also leads to a strong increase of the band broadening if supports with the same porosity ε and the same radial width are compared. For the presently considered 2D system, the flow resistance reduction is too small to overcome the corresponding strong increase in band broadening, such that the presence of preferential flow paths always leads to an overall increase of the separation impedance. © 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:53 / 61
页数:9
相关论文
共 26 条
[1]   Determination of the porosities of monolithic columns by inverse size-exclusion chromatography [J].
Al-Bokari, M ;
Cherrak, D ;
Guiochon, G .
JOURNAL OF CHROMATOGRAPHY A, 2002, 975 (02) :275-284
[2]  
[Anonymous], 1965, DYNAMICS CHROMATOGRA
[3]   STANDARDIZATION OF TEST CONDITIONS FOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY COLUMNS [J].
BRISTOW, PA ;
KNOX, JH .
CHROMATOGRAPHIA, 1977, 10 (06) :279-289
[4]   Influence of the pillar shape on the band broadening and the separation impedance of perfectly ordered 2-D porous chromatographic media [J].
De Smet, J ;
Gzil, P ;
Vervoort, N ;
Verelst, H ;
Baron, GV ;
Desmet, G .
ANALYTICAL CHEMISTRY, 2004, 76 (13) :3716-3726
[5]   ADVANCES IN BAND-SPREADING THEORIES [J].
GRUSHKA, E ;
KNOX, JH ;
SNYDER, LR .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1975, 13 (01) :25-37
[6]  
GUICHON G, 1994, FUNDAMENTALS PREPARA
[7]   A computational study of the porosity effects in silica monolithic columns [J].
Gzil, P ;
Vervoort, N ;
Baron, GV ;
Desmet, G .
JOURNAL OF SEPARATION SCIENCE, 2004, 27 (10-11) :887-896
[8]   Computational study of the band broadening in two-dimensional etched packed bed columns for on-chip high-performance liquid chromatography [J].
Gzil, P ;
De Smet, J ;
Vervoort, N ;
Verelst, H ;
Baron, GV ;
Desmet, G .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1030 (1-2) :53-62
[9]   Advantages of perfectly ordered 2-D porous pillar arrays over packed bed columns for LC separations: A theoretical analysis [J].
Gzil, P ;
Vervoort, N ;
Baron, GV ;
Desmet, G .
ANALYTICAL CHEMISTRY, 2003, 75 (22) :6244-6250
[10]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF PROTEINS ON COMPRESSED, NON-POROUS AGAROSE BEADS .1. HYDROPHOBIC-INTERACTION CHROMATOGRAPHY [J].
HJERTEN, S ;
LIAO, JL .
JOURNAL OF CHROMATOGRAPHY, 1988, 457 :165-174