Testing the reliability of a new ultrasonic cardiac output monitor, the USCOM, by using aortic flowprobes in anesthetized dogs

被引:66
作者
Critchley, LA [1 ]
Peng, ZY
Fok, BS
Lee, A
Phillips, RA
机构
[1] Chinese Univ Hong Kong, Prince Wales Hosp, Dept Anesthesia & Intens Care, Shatin, Hong Kong, Peoples R China
[2] Univ Queensland, Sch Med, Brisbane, Qld, Australia
关键词
D O I
10.1213/01.ANE.0000144774.42408.05
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We have used an animal model to test the reliability of a new portable continuous-wave Doppler ultrasonic cardiac output monitor, the USCOM. In six anesthetized dogs, cardiac output was measured with a high-precision transit time ultrasonic flowprobe placed on the ascending aorta. The dogs' cardiac output was increased with a dopamine infusion (0-15 mug (.) kg(-1) (.) min(-1)). Simultaneous flowprobe and USCOM cardiac output measurements were made. Up to 64 pairs of readings were collected from each dog. Data were compared by using the Bland and Altman plot method and Lin's concordance correlation coefficient. A total of 319 sets of paired readings were collected. The mean (+/-SD) cardiac output was 2.62 +/- 1.04 L/min, and readings ranged from 0.79 to 5.73 L/min. The mean bias between the 2 sets of readings was -0.01 L/min, with limits of agreement (95% confidence intervals) of -0.34 to 0.31 L/min. This represents a 13% error. In five of six dogs, there was a high degree of concordance, or agreement, between the 2 methods, with coefficients >0.9. The USCOM provided reliable measurements of cardiac output over a wide range of values. Clinical trials are needed to validate the device in humans.
引用
收藏
页码:748 / 753
页数:6
相关论文
共 19 条
  • [1] EVALUATION OF A TRANSIT-TIME SYSTEM FOR THE CHRONIC MEASUREMENT OF BLOOD-FLOW IN CONSCIOUS SHEEP
    BEDNARIK, JA
    MAY, CN
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1995, 78 (02) : 524 - 530
  • [2] Measuring agreement in method comparison studies
    Bland, JM
    Altman, DG
    [J]. STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) : 135 - 160
  • [3] THORACIC ELECTRICAL BIOIMPEDANCE MEASUREMENT OF CARDIAC-OUTPUT - NOT READY FOR PRIME-TIME
    CLARKE, DE
    RAFFIN, TA
    [J]. CRITICAL CARE MEDICINE, 1993, 21 (08) : 1111 - 1112
  • [4] The effectiveness of right heart catheterization in the initial care of critically ill patients
    Connors, AF
    Speroff, T
    Dawson, NV
    Thomas, C
    Harrell, FE
    Wagner, D
    Desbiens, N
    Goldman, L
    Wu, AW
    Califf, RM
    Fulkerson, WJ
    Vidaillet, H
    Broste, S
    Bellamy, P
    Lynn, J
    Knaus, WA
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (11): : 889 - 897
  • [5] A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques
    Critchley, LAH
    Critchley, JAJH
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) : 85 - 91
  • [6] Validation study of a new transit time ultrasonic flow probe for continuous great vessel measurements
    Dean, DA
    Jia, CX
    Cabreriza, SE
    DAlessandro, DA
    Dickstein, ML
    Sardo, MJ
    Chalik, N
    Spotnitz, HM
    [J]. ASAIO JOURNAL, 1996, 42 (05) : M671 - M676
  • [7] The reliability of vital sign measurements
    Edmonds, ZV
    Mower, WR
    Lovato, LM
    Lomeli, R
    [J]. ANNALS OF EMERGENCY MEDICINE, 2002, 39 (03) : 233 - 237
  • [8] NON-INVASIVE DOPPLER DETERMINATION OF CARDIAC-OUTPUT IN MAN - CLINICAL VALIDATION
    HUNTSMAN, LL
    STEWART, DK
    BARNES, SR
    FRANKLIN, SB
    COLOCOUSIS, JS
    HESSEL, EA
    [J]. CIRCULATION, 1983, 67 (03) : 593 - 602
  • [9] A CONCORDANCE CORRELATION-COEFFICIENT TO EVALUATE REPRODUCIBILITY
    LIN, LI
    [J]. BIOMETRICS, 1989, 45 (01) : 255 - 268
  • [10] Comparing methods of clinical measurement: Reporting standards for Bland and Altman analysis
    Mantha, S
    Roizen, MF
    Fleisher, LA
    Thisted, R
    Foss, J
    [J]. ANESTHESIA AND ANALGESIA, 2000, 90 (03) : 593 - 602