Dosimetric verification of open asymmetric photon fields calculated with a treatment planning system based on dose-to-energy-fluence concepts

被引:12
作者
Hurkmans, C
Knoos, T
Nilsson, P
机构
[1] Department of Radiation Physics, University Hospital in Lund
关键词
D O I
10.1088/0031-9155/41/8/003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Output normalized dose profiles for asymmetric open photon fields have been calculated using a commercial treatment planning system (TPS) based on a dose-to-energy-fluence concept. The model does not require any additional measurements for off-axis fields. Calculations are compared with measurements for quadratic fields of 5 cm x 5 cm up to 20 cm x 20 cm, with their geometric field centre positioned 10 cm off-axis in the in-plane direction. The measurements include depth doses and profiles in-plane as well as cross-plane for nominal photon energies of 4, 6 and 18 MV x-rays. Both calculated and measured doses are normalized with respect to a 10 cm x 10 cm reference field, therefore making it possible to compare not only the relative distributions but also the absolute dose levels; that is, calculation of monitor units is included. The calculated depth-dose curves are generally in good agreement with measured data with an accuracy at the absolute dose level of 2% at depths beyond the dose maximum. The cross-plane profiles are calculated with an accuracy better than 3% within the field. The 'tilt' towards the collimator central axis of the in-plane profiles is predicted by the model, but is somewhat overestimated at large depths. The system provides the possibility to separate the primary and scattered parts of the dose and the cause of this tilting was studied by comparing calculated phantom-scattering and head-scattering dose profiles for a symmetric 40 cm x 20 cm field to dose profiles for an asymmetric 20 cm x 20 cm field. The tilting is shown to originate from a change both in phantom scattering and in head scattering compared to the case of symmetrical fields. The results indicate that the investigated TPS can calculate dose distributions in open asymmetric fields with a high degree of accuracy, typically better than 2-3%.
引用
收藏
页码:1277 / 1290
页数:14
相关论文
共 24 条
[1]   ACQUISITION OF THE EFFECTIVE LATERAL ENERGY FLUENCE DISTRIBUTION FOR PHOTON-BEAM DOSE CALCULATIONS BY CONVOLUTION MODELS [J].
AHNESJO, A ;
TREPP, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (07) :973-985
[2]   A PENCIL BEAM MODEL FOR PHOTON DOSE CALCULATION [J].
AHNESJO, A ;
SAXNER, M ;
TREPP, A .
MEDICAL PHYSICS, 1992, 19 (02) :263-273
[3]   COLLIMATOR SCATTER IN PHOTON THERAPY BEAMS [J].
AHNESJO, A .
MEDICAL PHYSICS, 1995, 22 (03) :267-278
[4]   ANALYTIC MODELING OF PHOTON SCATTER FROM FLATTENING FILTERS IN PHOTON THERAPY BEAMS [J].
AHNESJO, A .
MEDICAL PHYSICS, 1994, 21 (08) :1227-1235
[5]   APPLICATION OF THE CONVOLUTION METHOD FOR CALCULATION OF OUTPUT FACTORS FOR THERAPY PHOTON BEAMS [J].
AHNESJO, A ;
KNOOS, T ;
MONTELIUS, A .
MEDICAL PHYSICS, 1992, 19 (02) :295-301
[6]   DETERMINATION OF EFFECTIVE BREMSSTRAHLUNG SPECTRA AND ELECTRON CONTAMINATION FOR PHOTON DOSE CALCULATIONS [J].
AHNESJO, A ;
ANDREO, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (10) :1451-1464
[7]  
BJARNGARD BE, 1994, MED PHYS, V21, P1069, DOI 10.1118/1.597349
[8]  
BRAHME A, 1988, ACTA ONCOL S, V1
[9]   MONTE-CARLO STUDY OF ACCELERATOR HEAD SCATTER [J].
CHANEY, EL ;
CULLIP, TJ ;
GABRIEL, TA .
MEDICAL PHYSICS, 1994, 21 (09) :1383-1390
[10]   DOSE COMPUTATIONS FOR ASYMMETRIC FIELDS DEFINED BY INDEPENDENT JAWS [J].
CHUI, CS ;
MOHAN, R ;
FONTENLA, D .
MEDICAL PHYSICS, 1988, 15 (01) :92-95