Portable automatic BTX measurement system with microfluidic device using mesoporous silicate adsorbent with nano-sized pores

被引:28
作者
Ueno, Y
Horiuchi, T
Niwa, O
Zhou, HS
Yamada, T
Honma, I
机构
[1] Natl Inst Adv Ind Sci & Technol, Tsukuba, Ibaraki 3058568, Japan
[2] NTT Microsyst Integrat Labs, Atsugi, Kanagawa 2430198, Japan
来源
SENSORS AND ACTUATORS B-CHEMICAL | 2003年 / 95卷 / 1-3期
关键词
microfluidic device; BTX gases; automatic measurement; portable system; mesoporous silicate;
D O I
10.1016/S0925-4005(03)00540-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We have developed a portable automatic measurement system for the detection and identification of aromatic volatile organic compound (VOC) gases, namely benzene, toluene, and xylenes (BTX), which are air pollutants, by assembling two microfluidic devices, a concentration cell and a detection cell, and peripheral devices. We combined the processes of gas concentration and thermal desorption performed by the concentration cell, with the process of spectroscopic detection performed by the detection cell. We achieved a 50 ppb detection limit for toluene gas as an example of BTX detection in a sampling time of 30 min. We also found that we could obtain better BTX separation by using a highly ordered mesoporous silicate as a gas adsorbent due to a function of its nano-sized pores. As a result, we successfully measured ppm level concentrations of BTX in a mixture gas separately with this device. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:282 / 286
页数:5
相关论文
共 5 条
[1]  
[Anonymous], 1999, Compendium of Methods for the Determination of Toxic Organic Compounds in Ambient Air, V2nd ed.
[2]   Air-cooled cold trap channel integrated in a microfluidic device for monitoring airborne BTEX with an improved detection limit [J].
Ueno, Y ;
Horiuchi, T ;
Niwa, O .
ANALYTICAL CHEMISTRY, 2002, 74 (07) :1712-1717
[3]   Separate detection of BTX mixture gas by a microfluidic device using a function of nanosized pores of mesoporous silica adsorbent [J].
Ueno, Y ;
Horiuchi, T ;
Tomita, M ;
Niwa, O ;
Zhou, HS ;
Yamada, T ;
Honma, I .
ANALYTICAL CHEMISTRY, 2002, 74 (20) :5257-5262
[4]   Microfluidic device for airborne BTEX detection [J].
Ueno, Y ;
Horiuchi, T ;
Morimoto, T ;
Niwa, O .
ANALYTICAL CHEMISTRY, 2001, 73 (19) :4688-4693
[5]   Nonionic triblock and star diblock copolymer and oligomeric surfactant syntheses of highly ordered, hydrothermally stable, mesoporous silica structures [J].
Zhao, DY ;
Huo, QS ;
Feng, JL ;
Chmelka, BF ;
Stucky, GD .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1998, 120 (24) :6024-6036